Oxygen analyser (ppb) - 9182 -
6-2
6.1.1 Automatic temperature compensation
The sensor measures continuously the sample temperature. The concentration values are
automatically calculated according to the reference temperature (25 °C) by a preprogrammed
compensation law in the transmitter.
Follow the procedure below:
PROGRAMMING
Choice of an
automatic
temperature
compensation.
EXECUTION
The execution is realised with the
CALIBRATION menu.
Enter the sample temperature value in °C.
Press OK to adjust the temperature displayed
to the actual value of the sample temperature
measured with a precise thermometer.
The gap between the calibration and the
theorical response curve of the sensor is given
for information.
MEASURE
Select
TEMP. COMP.
TYPE :
SENSOR : NTC
TEMP : 25.0°C
TEMP. COMP.
Select
Enter
Auto.
TEMP. CALIB.
Select
EXECUTION
21.2
°C
6.87 ppm
OK
CAL
21.2
°C
1.2 °C
CAL
2
Δ
T : 0.0 °C
ENTER : ok
ESC : cancel
TEMP. CALIB.
Enter
Enter
Press OK when the
measure is stable
Possibility to change
the value
Summary of Contents for POLYMETRON 9182
Page 1: ...Operator Manual POLYMETRON 9182 Oxygen analyzer ppb 221 191 082 Revision G 20 11 2007 ...
Page 2: ......
Page 8: ......
Page 39: ...Oxygen analyser ppb 9182 5 9 ...
Page 48: ......
Page 56: ......
Page 70: ......
Page 74: ......
Page 76: ......
Page 80: ......
Page 86: ......
Page 90: ...Oxygen analyser ppb 9182 A7 4 ...
Page 91: ......
Page 92: ......