Oxygen analyser (ppb) - 9182 -
8-2
CURRENT
TOO LOW
The current value is out of the limits. Check the
electrode (electrolyte and membrane).
CURRENT
TOO HIGH
The current value is out of the limits. Check there is no
short-circuit on the measuring line.
Check the polarization voltage.
Error messages during a calibration
∆
T
OUT OF LIMITS
21.6
°C
The temperature difference between the calibration and
the sensor theorical response is superior to the
programmed limit.
Limits:
±
20
°
C
Summary of Contents for POLYMETRON 9182
Page 1: ...Operator Manual POLYMETRON 9182 Oxygen analyzer ppb 221 191 082 Revision G 20 11 2007 ...
Page 2: ......
Page 8: ......
Page 39: ...Oxygen analyser ppb 9182 5 9 ...
Page 48: ......
Page 56: ......
Page 70: ......
Page 74: ......
Page 76: ......
Page 80: ......
Page 86: ......
Page 90: ...Oxygen analyser ppb 9182 A7 4 ...
Page 91: ......
Page 92: ......