Oxygen analyser (ppb) - 9182 -
7-6
7.2.2 Mounting the new membrane
F.
Screw the new membrane on the body till the dead stop.
G.
Replenish the probe body with 5 ml of electrolyte. Check there is no impurity or bubble in the
electrolyte.
H.
Preposition the electrode without forcing. It should take its place by simple gravity.
I.
Screw the probe nut till the electrode meets a resistance.
J.
Put back the cap and gasket, screw the cap to ensure watertightness.