Oxygen analyser (ppb) - 9182 -
7-12
MALFUNCTION
POSSIBLE CAUSE
REMEDIES
Measurement is not
correct
Instrument was programmed
incorrectly
Recheck programmed
parameters. Do they agree with
the probe's characteristics?
System, including probe, not
calibrated correctly
Calibrate the whole system
(probe connected)
Probe connected wrong
Recheck all probe connections
Probe malfunctioning,
possibly incompatible with the
application
Visually check the condition of
the probe. Is the application
within the probe's
specifications (data sheet)?
CPU board is defective
If error persists, call the
Service Technician
Measurement is not
stable
Faulty probe
Check the condition of the
probe. Is it contaminated?
Probe connected wrong
Is the probe connected
correctly?
Interferences
Are there any sources of
potential interference,
chemical, external,
temperature, pressure, ...?
Cable shield is not connected
Check and connect
Defective CPU board
If problem persists, call the
Service Technician
Probe connected wrong
Is the probe properly
connected? Check
Temperature
measurement is not
correct
Temperature was not
calibrated
Calibrate for temperature. Also
check the Pt100 for correct
resistance (ohmmeter)
CPU board is defective
If problem persists, call the
Service Technician
Summary of Contents for POLYMETRON 9182
Page 1: ...Operator Manual POLYMETRON 9182 Oxygen analyzer ppb 221 191 082 Revision G 20 11 2007 ...
Page 2: ......
Page 8: ......
Page 39: ...Oxygen analyser ppb 9182 5 9 ...
Page 48: ......
Page 56: ......
Page 70: ......
Page 74: ......
Page 76: ......
Page 80: ......
Page 86: ......
Page 90: ...Oxygen analyser ppb 9182 A7 4 ...
Page 91: ......
Page 92: ......