Options - 3625 Package Analyzer
Operator Manual
ORBISPHERE
Procedure
•
ESC
displays main menu
•
×
Ø
select
OPTIONS
;
ENTER
•
×
Ø
select
SERIAL OUTPUT
;
ENTER
•
×
Ø
select
DIAGNOSTIC
;
ENTER
•
×
Ø
select
CORRECTIONS
;
ENTER
•
×
Ø
select
HUMIDITY
;
ENTER
•
×
Ø
select
CALIBRATE
;
ENTER
• Wait for stabilization of slope value.
• Check value slope on the
Diagnostic View
window of the PC.
• When slope is stable (-1 [mV/sec] ±1), press
ENTER
to save the new humidity
correction factor.
6.5.2 Oxygen Factor Correction
Depending on concentration, O
2
in the measured gas may affect the N
2
sensor readings.
The following section concerns setting the oxygen correction factor of the N
2
sensor.
Note:
Calibrate both O
2
and N
2
sensors before performing this.
Procedure
• Expose the N
2
sensor to the atmosphere, beside the O
2
sensor.
• On PC, start the OrbiPack software.
• Display
Diagnostic view
by selecting it in the
Tools
menu.
On measuring instrument:
•
×
Ø
select
OPTIONS
;
ENTER
•
×
Ø
select
CORRECTIONS
;
ENTER
•
×
Ø
select
OXYGEN
;
ENTER
•
×
Ø
select
CALIBRATE
;
ENTER
Summary of Contents for ORBISPHERE 3625
Page 1: ...Operator Manual ORBISPHERE 3625 Revision F 04 04 2008 ...
Page 2: ......
Page 14: ...8 of 90 Manual Overview 3625 Package Analyzer Operator Manual ORBISPHERE ...
Page 52: ...46 of 90 Automatic Analysis Sequence 3625 Package Analyzer Operator Manual ORBISPHERE ...
Page 60: ...54 of 90 Operating Instructions 3625 Package Analyzer Operator Manual ORBISPHERE ...
Page 82: ...76 of 90 Maintenance 3625 Package Analyzer Operator Manual ORBISPHERE ...
Page 92: ...86 of 90 Glossary 3625 Package Analyzer Operator Manual ORBISPHERE ...
Page 94: ...88 of 90 Annex 3625 Package Analyzer Operator Manual ORBISPHERE ...
Page 95: ...3625 Package Analyzer User Notes 89 of 90 Operator Manual ORBISPHERE User Notes ...
Page 96: ......