background image

 

file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc 

© 2015 h/p/cosmos sports & medical gmbh     [email protected]     created 30.01.2015     printed 30.01.2015      page: 1 of 216 

 

 

 

 

original instructions for use 

 

h/p/cosmos

®

 running machines 

 

N

 

150/50 LC (pluto

®

N

 

150/50 (mercury

®

N

 

170-190/65 (quasar

®

 / pulsar

®

N

 

190/65 3p (pulsar

®

 3p) 

N

 

200-300/75-125 (venus

®

 / saturn

®

N

 

450/300 (saturn

®

h/p/cosmos

®

 model names 

pluto

® 

/ pluto

® 

med 

mercury

®

 / mercury

®

 med 

stratos

® 

/ stratos

®

 med 

locomotion

®

 150/50 E / DE med 

gaitway

®

 II F / S 

incl. lt-models 
 
quasar

®

 / quasar

®

 med 

stellar

®

 / stellar

®

 med 

pulsar

®

 

locomotion

®

 190/65 E / DE med 

incl. lt-models 
 
quasar

®

 med 3p 

pulsar

®

 3p 

locomotion

®

 190/65 3p E / DE med 

incl. lt-models 
 
venus

®

 - all sizes 

saturn

®

 - all sizes 

 

 

development, production, sales 

&

 service 

h/p/cosmos

®

 sports 

&

 medical gmbh 

Am Sportplatz 8 
DE 83365 Nussdorf-Traunstein 
Germany 
phone +49 86 69 86 42 0 
fax +49 86 69 86 42 49 
[email protected] 
www.h-p-cosmos.com 
 
 

manual-version 

revision 29.01.2015 
firmware MCU5-v1.08 
manual order-no.: [cos14310m5-v1.08hpc-en] 

Summary of Contents for 190/65 3p

Page 1: ...osmos model names pluto pluto med mercury mercury med stratos stratos med locomotion 150 50 E DE med gaitway II F S incl lt models quasar quasar med stellar stellar med pulsar locomotion 190 65 E DE med incl lt models quasar med 3p pulsar 3p locomotion 190 65 3p E DE med incl lt models venus all sizes saturn all sizes development production sales service h p cosmos sports medical gmbh Am Sportplat...

Page 2: ... fitness performance diagnostics but is not sufficient for all special high performance applications So the 3 phase running machine models with 3 x 380 420 volts power supply for example model pulsar 3p venus or saturn are the perfect and recommended solutions for high performance applications In case of single phase and or low voltage supply there may performance limitations and limitation of max...

Page 3: ...horized technicians The customer user should perform this maintenance and monitoring work Some safety checks or monitoring for examples harnesses and ropes running belt condition and position etc have to be performed on daily basis It is not expedient to contract certified technicians for such maintenance work However where it is practical all maintenance and monitoring work marked with this symbo...

Page 4: ... 3 1 Switch on procedure model range 150 50 LC pluto 28 5 3 2 Switch on procedure model range 150 50 up to 190 65 mercury quasar pulsar 29 5 3 3 Switch on procedure model range 200 75 up to 300 125 venus saturn 30 5 3 4 Switch on procedure model saturn 450 300 31 5 4 Switch off procedure 33 5 4 1 Switch off procedure model range 150 50 LC pluto 33 5 4 2 Switch off procedure model range 150 50 up t...

Page 5: ... notes warnings precautions 84 7 1 1 Room conditions 84 7 1 2 Transport and Installation 85 7 1 3 Maintenance and safety inspections 86 7 2 Electrical safety features 87 7 3 Medical device software classification 88 7 4 Installation commissioning 89 7 4 1 Unpacking and packaging 89 7 4 2 Transport 90 7 4 3 Storage 93 7 4 4 Environmental requirements room configuration 94 7 4 5 Mechanical installat...

Page 6: ...ance RPE measurement 116 8 3 3 Isolation resistance RISO measurement 116 8 3 4 Equivalent alternative leakage current IEDL measurement 116 8 3 5 Leakage current measurement 117 8 3 6 Electric safety tester and measurements 117 8 3 7 Building installation Electric checks protective earth function RCD leakage current 118 8 4 Lubrication of the running belt running surface 118 8 4 1 Devices with User...

Page 7: ... Hygiene and cleansing 132 8 11 1 Cleansing of outside and applied parts 132 8 11 2 Cleaning of the interior 133 8 12 Cleansing and adjustment of the speed sensor light barrier 133 8 13 Light barrier at belt re entry zones 134 8 13 1 1 Cleaning of light barriers 134 8 13 2 Adjustment of light barriers 134 8 14 Spare parts consumables 134 9 Trouble shooting 135 9 1 Device cannot be powered on 135 9...

Page 8: ...r tooth locking elements 177 12 5 Handrails adjustable Parallel bars with gas struts and locking lever 178 12 6 Therapist s seat and foot support adjustable 180 12 7 Unweighting system h p cosmos airwalk se 135 181 12 8 Unweighting systems airwalk 70 160 182 12 9 Vests for subjects for airwalk systems 182 12 10 h p cosmos robowalk expander 182 12 11 Wheelchair ramp 183 12 12 Wheelchair stabilizer ...

Page 9: ...he TSA TÜV according to EN ISO 13485 201 14 3 EC Certificate of the TSA TÜV 202 15 Appendix 2 203 15 1 Calculation of the UKK Fitness Index 203 15 2 Instruction protocol checklist 204 15 3 Instruction protocol signatures 209 15 4 Service report cos15531 210 15 5 Control protocol cos11690en 02 211 15 6 Wiring diagram load circuit quasar med pulsar 212 15 7 Wiring diagram control circuit quasar med ...

Page 10: ...h p cosmos sports fitness running machines pluto med pluto lt med pluto pluto lt locomotion 150 50 E med locomotion 150 50 DE med locomotion 190 65 E med locomotion 190 65 DE med locomotion 190 65 3p E med locomotion 190 65 3p DE med stratos lt med stratos med mercury lt med mercury med gaitway II F gaitway II S stratos lt stratos mercury lt mercury stellar lt med stellar med quasar lt med quasar ...

Page 11: ... medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 11 of 216 venus 200 75 venus 200 75r venus 200 100 venus 200 100r saturn 250 75 saturn 250 75r saturn 250 100 saturn 250 100r saturn 250 125r saturn 300 75 saturn 300 75r saturn 300 100 saturn 300 100r saturn 300 125r saturn 450 300rs saturn 300 125r saturn 450 300rs ...

Page 12: ...y secured and un weighted with a body weight support system and the gait can be analysed and corrected under safe and reproducible conditions Duration of use of the medical versions is approx 5 to 25 minutes per patient per day depending on the prescription of the doctor the condition of the patient and depending on the standardized or individual load protocol for example BRUCE protocol The runnin...

Page 13: ...te control of the treadmill loading parameters and for maintenance issues Software products such as h p cosmos para graphics h p cosmos para analysis and h p cosmos para motion can be optionally applied for further data analysis and documentation especially to be used for sports and fitness applications The h p cosmos software is neither intended for medical diagnosis nor medical evaluation The me...

Page 14: ...chines It can for example be duplicated on the opposite handrail h p cosmos conforms with the norm EN 957 6 The push button type or pull cord safety stop actuator shall be in a reachable position in front of the user and in an area of 180 mm parallel to the centre line of the running surface If the push button type safety stop actuator is outside this area it shall be duplicated on the opposite si...

Page 15: ...t not be operated in home environment The correct loading for a patient on the treadmill must be prescribed by a medical doctor The treadmill manufacturer cannot make any declaration or recommendation The following table gives an overview over the intended applications and the corresponding minimum safety equipment A not restricted patient is defined as a patient able to grap the handrail and jump...

Page 16: ...on N Acute endocarditis myocarditis pericarditis N Acute aortic dissection N Acute coronary syndrome N Acute phlebothrombosis of the lower extremities N Febrile infections N Pregnancy N Acute thrombosis N Fresh wounds e g after surgery N Acute fracture N Damaged disc or traumatic disease of the spine N Epilepsy N Inflammations N Acute migraine Relative contraindications N Left main coronary stenos...

Page 17: ...gnition restrictions be excluded by contraindications The risk of falling and consequential injuries is reduced by the safety equipment and application of a fall prevention system Remaining risks in ergometry stress testing The following complications are mentioned in the Praxisleitlinien Ergometrie Wonisch M Berent R Klicpera M Laimer H Marko C Pokan R Schmid P Schwann H Practice Guidelines for E...

Page 18: ... personnel must stay in permanent patient environment of 1 5 m N Recommendation and application of weight support system and or fall stop device such as the h p cosmos safety arch with harness and automatic shut off of the running belt N Comprehensive instruction of the patient before the start of treadmill therapy with an explanation that the drive of the running belt is very strong and does not ...

Page 19: ...contraindications prevail see chapter entitled Contraindications N The automatic operation of the running machine modes profile cardio test remote control via PC and peripheral devices is forbidden if the health and condition of the test person subject patient does not permit and if a doctor has not authorized said operation Disregarding this provision may lead to injuries serious health problems ...

Page 20: ... accessories without supervisory staff and introduction N Holding the handrails during exercise may affect the exercise results e g heart rate ECG oxygen uptake etc Under normal conditions and for healthy subjects we recommend holding the handrails only in case of emergency or when needed for safety or weight relief N Under no circumstances should the device be used with increased risk e g sprint ...

Page 21: ...ave an extra thick running belt A safety arch with chest belt system is obligatory in this case N In case of reverse belt rotation always use the safety arch with chest belt system since the standard safety stop buttons and keyboard as well as standard handrails and front bar handrail might not be in reach when facing the running deck backwards Furthermore a supervisor must be present 4 3 Preparat...

Page 22: ...e of severe danger N subjects with long hair or wide clothes N The safety arch with chest belt system must be checked for wear and damage before each use In particular the rope harness and all links such as the snap hook and the rope brake are to be checked N All wear and tear parts of the system rope harness and all links such as snap hook and rope brake are to be changed immediately in case of d...

Page 23: ...io mode in case of detected or suspected interference of the wireless heart rate transmission N The subject must interrupt training immediately if he she starts feeling sick dizzy or feels pain The patient should consult his her doctor in this case N If the treadmill is being controlled via ECG pay special attention to the ECG manufacturer s manual The summary of clinical data and the risk managem...

Page 24: ...p cosmos strongly recommends keeping the standard setting of this option Stop means stop for all movements and should not activate the running machine in any way It is only allowed to change option 41 after having received written instructions regarding the dangers and after written confirmation of receipt by the customer user N Motor break Attention If the device is switched off or disconnected f...

Page 25: ...ty arch with chest belt system and stop function unweighting systems should be adjusted so that the subject cannot use the rear 30 of the running deck length and the automatic stop function stops the running belt The patient subject is placed onto the correct position on the treadmill Before using the treadmill the patient subject has to be carefully instructed to follow the speed of the running b...

Page 26: ...n 450 300 is the same as at the regular smaller h p cosmos treadmills Controversial to the smaller treadmills the elevation element is powered by a hydraulic system Illustration Description Cooler Electronic valve Face valve Hydraulic oil tank Manometer Tank lid Temperature sensor Visual display for oil level and oil temperature On the saturn 450 300 the safety arch is movable and can be adjusted ...

Page 27: ...e belt fits tightly but is not confining The belt should not loosen while exercising Close the belt with the transmitter placed outwards POLAR logo in right position In order to allow optimal skin contact the skin should be damp Contact gel as used for ECG is an excellent moisturizer Moisten the two electrodes and the skin with water or ECG contact gel Place the transmitter so that it is directly ...

Page 28: ... on the running machine by using the silver push button at the UserTerminal The indicator light within the push button is illuminated blue If the indicator does not flash up please check the power supply the device protection switch and the emergency stop See chapter 9 1 Device cannot be powered on N If the running machine is at an angle of elevation 0 when switched on it will automatically move t...

Page 29: ...e sure that the expulsion fuse at the front of the device front terminal below the hood is switched on N If the running machine is at an angle of elevation 0 when switched on it will automatically move to the position zero display elevation INIT Please ensure that while lowering the elevation no harm is done to persons or objects under the running machine N Plug the running machines directly into ...

Page 30: ...appears on the display Expulsion fuse and main switch 2 or 3 pole ON OFF button with control lamp ON OFF button until 2003 5 3 3 Switch on procedure model range 200 75 up to 300 125 venus saturn no Illustration Description 01 Plug the power cord of the running machine directly into the CEE wall socket Only the model saturn 450 300rs has a permanent connection to the mains and no power plug N If th...

Page 31: ...hnicians 04 Switch on the running machine by using the main switch at the right side of the external UserTerminal control unit The main switch has to be turned 90 clockwise as shown in the picture The main switch can be secured against unauthorized access with a padlock 05 After a delay time green light is blinking of 1 minute the green light stops blinking and stays illuminated Now the running ma...

Page 32: ... of the device must only be performed by authorized technicians 03 Switch on the running machine by using the main switch at the right side of the external UserTerminal control unit The main switch has to be turned 90 clockwise as shown in the picture The main switch can be secured against unauthorized access with a padlock 04 Press the power on push button After a delay time green light is blinki...

Page 33: ...ption 02 Switch off the running machine by pressing the black or red OFF push button at the front of the device The indicator light within the key extinguishes For some models the location of the ON OFF key may also be at the UserTerminal Devices for sports and fitness manufactured before 12 2007 can be switched off directly with the expulsion fuse at the FrontTerminal frontal section below the ho...

Page 34: ...or medical application with the isolation transformer have an inrush current limiter surge guard Too short ON OFF intervals will lead to a deactivation of the inrush current limiter and result in an overload of the circuit fuse N If the device is used daily we recommend switching it on in the morning and leave it in stand by mode during the day N Press the emergency stop only in case of danger N T...

Page 35: ...l or new command via interface Quick Stop models venus saturn Emergency stop button is connected to the quick stop port of the UserTerminal N mushroom switch is hit N running belt inclination system stop extremely fast N drive motor is used to stop the running belt very fast N machine is NOT cut off from main power of drive motor and inclination system N display active shows PULL STOP N interface ...

Page 36: ...erminal N press START key or new command via interface Additional stop button for arm support N Stop button is pushed N running belt inclination system stop fast N machine is NOT cut off from main power of drive motor and inclination system N display active shows PULL STOP N interface active but no start speed or elevation signal executed N keyboard inactive NOT interrupted during PULL STOP stage ...

Page 37: ...minal control desk UserTerminal TouchPanel retro fit kit If your running machine has no UserTerminal no display no keyboard it can only be controlled via interface RS 232 e g via ECG Ergospirometry PC with software h p cosmos para graphics or h p cosmos para control Compatible peripheral devices are listed in chapter 12 entitled Accessories and options More recent computers will only have a USB in...

Page 38: ...f six four digit LCD displays which show speed time index MET energy consumption wattage alternating distance elevation inclination and heart rate Error codes and service information can also be displayed Next to the LCD displays you will find light emitting diodes LED which indicate the selected operation mode or measuring unit LED Flashing displays indicate N select option LED N alterable parame...

Page 39: ...ff speed LED status unit definition max m min km h m s mph m min km h m s mph 0 1 0 1 or 0 01 0 1 or 0 01 0 1 or 0 01 running distance LED status unit definition m km miles m km miles 0 1 0 1 0 1 running time LED status unit definition mm ss hh mm 1 1 elevation program sequence step number no LED status unit definition step no 0 1 0 1 1 1 display LED status unit definition MET energy consumption w...

Page 40: ...ckward switch is blue lighted belt alignment adjust it Running belt is moving outside tolerance range yellow lighted belt out of tolerance yellow blinking belt severe out of tolerance check power supply Power supply too high or low or phase breakdown Check only by trained and authorized technician oil tank empty Refill oil tank see chapter 8 4 5 light barrier quick stop active One or more safety f...

Page 41: ...urface of the screen The TouchPanel PC turns on automatically with the external control unit In case the TouchPanel PC does not boot up it can be switched on manually Therefore the external control unit has to be opened in order to reach the power switch at the rear of the TouchPanel PC Opening of the device must only be performed by authorized technicians After booting up h p cosmos para control ...

Page 42: ...e speed display next target elevation with a click on the elevation display direct change of modes cool down quick stop count down option settings etc For details read the manual of h p cosmos para control http www h p cosmos com downloads manual 20110711_cos10071 v4 0man en_manual_h p cosmos_para_control_4 0 pdf The red coloured STOP button next to the START button on the UserTerminal is NOT an e...

Page 43: ... the power switch at the rear of the TouchPanel PC Opening of the device must only be performed by authorized technicians Before the running machine can be used the safety instructions have to be read and confirmed with a click on the respective confirm button I fully understand and accept Furthermore there are a lot of additional functions like next target speed with a click on the speed display ...

Page 44: ...f both The protocols can be selected in the test mode see chapter 5 10 5 Test mode You can also create a user defined profile either using the graded test or the freely definable user profiles Also see guidelines of German or American Heart Associations Deutsche Gesellschaft für Kardiologie Herz und Kreislaufforschung e V http leitlinien dgk org American Heart Association http www heart org and ht...

Page 45: ...nd use it as basis for therapy Gait analysis is indicated in the following cases N Pathologic gait pattern arthritis in ankles N Femoropatellar syndrome N Knee joint problems N Anterior cruciate ligament reconstruction The following figures show four examples of the clinical benefit of h p cosmos gaitway with KISTLER force plates Arthritis in the upper ankle 4 weeks later Femoropatellar syndrome 4...

Page 46: ... recommend using a printer linked to the serial interface RS 232 or and using the software h p cosmos para graphics on an external PC N The treadmill belt will start moving automatically and the speed and elevation will be changed automatically after an automatic mode profile cardio test or control via interface has been started All users need to be familiar with the details and risks of these mod...

Page 47: ...ofile mode and test mode The acceleration levels are accessible by pressing the speed keys and minus several times and then holding the key The acceleration levels are also accessible via interface and optional PC software h p cosmos para control and h p cosmos para graphics a speed in km h b time in seconds Example Time to a maximum speed of 40 km h Intensity level 1 131 sec 2 66 sec 3 33 sec 4 1...

Page 48: ...of activation in the optional functions indicates 65 flashing for 65 kg body weight weight is illuminated Enter the actual body weight with a or d then c for an accurate calculation of the POWER and ENERGY consumption Further steps see A 03 Alteration of speed aor d press and hold Speed is being increased reduced 0 max If speed is being reduced to 0 PAUS is displayed Current values will be sustain...

Page 49: ... the maximum speed maximum elevation and the duration of the six profiles can be scaled in order to provide 6 x 6 x 6 216 variations of profiles Standard set up is without scaling for quicker and easier access Profiles 1 3 are time dependent and without elevation profiles 4 6 are time dependent and with elevation As optional feature for some models on memory location 30 99 additional program profi...

Page 50: ... 3 Progressive interval m sec km h min Warming up 2 8 10 1 4 0 0 0 High 1 3 2 11 5 2 0 0 0 Low 1 2 8 10 1 2 0 0 0 High 2 3 6 13 0 1 0 0 0 Low 2 2 8 10 1 2 0 0 0 High 3 4 0 14 4 1 0 0 0 Low 3 2 8 10 1 2 0 0 0 High 4 3 6 13 0 1 0 0 0 Low 4 2 8 10 1 2 0 0 0 High 5 3 2 11 5 1 0 0 0 Low 5 2 8 10 1 2 0 0 0 High 6 3 2 11 5 1 0 0 0 Low 6 2 8 10 1 4 0 0 0 25 min Profile 4 Endurance interval for beginners m...

Page 51: ... 0 0 Low 1 2 0 7 2 2 0 10 0 High 2 2 5 9 0 3 0 0 0 Low 2 2 0 7 2 2 0 10 0 High 3 2 5 9 0 3 0 0 0 Low 3 2 0 7 2 2 0 10 0 High 4 2 5 9 0 3 0 0 0 Low 4 2 0 7 2 2 0 10 0 25 min Profile 6 Progressive interval elevation m sec km h min Warming up 2 8 10 1 4 0 0 0 High 1 3 2 11 5 2 0 10 0 Low 1 2 8 10 1 2 0 0 0 High 2 3 6 13 0 1 0 7 5 Low 2 2 8 10 1 2 0 0 0 High 3 4 0 14 4 1 0 5 0 Low 3 2 8 10 1 2 0 0 0 H...

Page 52: ... selected profile no 1 6 04 Start the selected profile no 1 6 c A standard in case of deactivation in the optional functions profile is illuminated and step are illuminated alternately kJ and Watt are illuminated alternately Running belt accelerates up to the starting speed of the first program step 5 seconds before the next program step there is an acoustic countdown Start of measuring All displa...

Page 53: ...ation c press several times and then hold Example to select acceleration or deceleration level 3 press the corresponding key 3 times and then hold Standard value 4 maximum 7 levels may be limited by optional settings 03 Pause mode interfere with a program step a indicates PAUS Running belt stops All displays stop Current values will be sustained After restart with d or c all displayed values will ...

Page 54: ...eath N If the user feels faint sick dizzy or pain or in case of any other health problems stop exercising immediately and see a doctor N In case of detected or suspected interference of the wireless heart rate transmission do not use the cardio mode N The operation of the cardio mode is prohibited if the health and the condition of the subject patient do not permit and a doctor has not authorized ...

Page 55: ...on The running machine adjusts speed and elevation according to the following matrix Chart for heart rate controlled work load at the heart rate lower level Heart rate difference actual set value Speed km h Elevation Time s 0 5 0 2 0 1 25 5 1 15 0 4 0 2 25 15 1 30 0 6 0 4 25 30 1 50 0 8 0 8 20 50 1 0 1 0 20 Chart for heart rate controlled work load at the heart rate upper level Heart rate differen...

Page 56: ...f your medical doctor you can must change this level If no changes are required confirm the value with c 07 Change upper level of heart rate aor d is illuminated flashing number indicates max level of the heart rate 08 Confirm upper level of heart rate c is illuminated flashing number indicates the suggested min level of the heart rate Calculated with the formula max level of heart rate minus 10 0...

Page 57: ...nt the subject to control the load by means of speed without elevation If speed is being reduced to 0 PAUS is displayed Current values will be sustained 02 Pause mode Interfere with the cardio mode a indicates PAUS running belt stops All displays stop Current values will be sustained After restart with d or c all displayed values will be added 03 Alteration of elevation for e press and hold Elevat...

Page 58: ...aded test COOPER test etc are endurance tests max load and max heart rate tests and should only be performed after consulting a medical doctor and under supervision of trained staff N When using the treadmill for load tests a medical doctor and a defibrillator must always be within reach N Proper warm up and cool down periods have to be considered N Use a safety arch with chest belt system when ma...

Page 59: ...vels limited by the maximum acceleration level selected in the optional settings standard value 4 During deceleration this function alternates the degree of deceleration 06 Switch to the next or previous mode c d or c a Switch to the next or previous mode This is a perfect way to change the mode without stopping the belt 5 10 5 1 Standard ergometric tests 1 UKK walk test The UKK walk test no 1 is ...

Page 60: ...ing 06 Select age 30 years aor d Age is illuminated indicates age 30 flashing 07 Confirm 30 years c Weight is illuminated indicates weight 65 flashing 08 Select weight 55 kg aor d Weight is illuminated indicates weight 55 flashing 09 Confirm 55 kg c indicates height H175 flashing 10 Select height 170 cm aor d indicates height H170 flashing 11 Confirm 170 cm c The running machine starts indicates s...

Page 61: ...on level 4 adjustable from 1 to 5 N Break time 30 sec can be changed STOP must be activated manually by the medical doctor In an already pre defined step test it is possible to set the pre programmed pause times individually during the test i e the pause can be prolonged or ended immediately End pause manually Once the test profile has reached the pause mode and the running belt has come to a halt...

Page 62: ...rding to the condition of the subject N Circuit lap length 200 m can be changed N Increment 0 5 km h can be changed STOP must be activated manually by the medical doctor when the subject is completely exhausted 4 Bruce protocol e g for ECG stress test Step Duration min Speed km h Elevation 1 3 00 2 7 10 2 4 0 12 3 5 4 14 4 6 7 16 5 8 0 18 6 8 8 20 7 9 6 22 5 Naughton protocol e g for ECG stress te...

Page 63: ... and the speed is increased by 0 32 km h every minute STOP must be activated manually by the medical doctor when the subject is completely exhausted 8 Ellestad A protocol e g for ECG stress test Step Duration min Speed km h Elevation 1 3 00 2 7 10 0 2 4 8 3 6 4 4 8 0 9 Ellestad B protocol e g for ECG stress test Step Duration min Speed km h Elevation 1 3 00 2 7 10 0 2 4 8 10 0 3 6 4 10 0 4 8 0 10 ...

Page 64: ...the displays will be reset to zero After completing the test the results can be printed on a host printer if connected Pre test phase patient stands on the footboards not on the belt Step Duration min sec Speed km h Elevation Total time min sec 0 untilSTART ispressed 3 2 0 untilSTART ispressed Test phase patient steps onto the running belt 1 02 00 3 2 0 2 00 2 02 00 3 2 2 4 00 3 02 00 3 2 4 6 00 4...

Page 65: ...n 0 0 9 500 m 19 0 10 1 15 min 0 0 72 Training Interval 400 m 16 0 21 0 km h Step Distance time Speed km h Elevation Acceleration level 1 400 m 16 0 0 3 2 1 00 min 0 0 3 400 m 17 0 4 1 00 min 0 0 5 400 m 18 0 6 1 00 min 0 0 7 400 m 19 0 8 1 00 min 0 0 9 400 m 20 0 10 1 00 min 0 0 11 400 m 21 0 12 1 00 min 0 0 73 Training Interval 300 600 m 15 0 21 0 km h pyramid Step Distance time Speed km h Eleva...

Page 66: ...m 15 0 0 3 2 1 30 min 0 0 3 300 m 21 0 4 0 45 min 0 0 5 600 m 15 0 6 1 30 min 0 0 7 300 m 21 0 8 0 45 min 0 0 9 600 m 15 0 10 1 30 min 0 0 11 300 m 21 0 12 0 45 min 0 0 13 600 m 15 0 14 1 30 min 0 0 15 300 m 21 0 16 0 45 min 0 0 75 Training Interval 300 m 20 0 km h Step Distance time Speed km h Elevation Acceleration level 1 300 m 20 0 0 3 2 0 45 min 0 0 3 300 m 20 0 4 0 45 min 0 0 5 300 m 20 0 6 ...

Page 67: ... 4 9 7 1 00 5 5 8 1 00 6 1 9 1 00 6 7 10 1 00 7 3 11 1 00 7 9 12 1 00 8 5 13 3 00 1 8 Tests no 77 79 are reserved for the following extensions and up dates 80 Profile VO2 10 k Step Time min Speed km h Elevation Acceleration level 1 2 00 8 0 0 1 2 2 00 10 0 0 3 1 00 10 0 2 4 1 00 10 0 4 5 1 00 10 0 6 6 1 00 10 0 8 7 1 00 10 0 10 8 1 00 11 0 10 9 2 00 11 0 11 10 2 00 11 0 12 81 Profile VO2 11 k Step...

Page 68: ...0 8 7 1 00 12 0 10 8 1 00 13 0 10 9 2 00 13 0 11 10 2 00 13 0 12 83 Not engaged reserved for following extensions and up dates 84 Profile VO2 14 k Step Time min Speed km h Elevation Acceleration level 1 2 00 12 0 0 1 2 2 00 14 0 0 3 1 00 14 0 2 4 1 00 14 0 4 5 1 00 14 0 6 6 1 00 14 0 8 7 1 00 14 0 10 8 1 00 15 0 10 9 2 00 15 0 11 10 2 00 15 0 12 85 Profile Super Balke Step Time min Speed km h Elev...

Page 69: ...tion Acceleration level 1 3 00 7 5 0 1 2 7 5 3 3 7 5 6 4 8 0 6 5 8 0 9 6 8 0 12 7 8 0 15 8 8 0 18 9 8 0 21 10 8 0 24 91 Test profile Step Time min Speed km h Elevation Acceleration level 1 3 00 9 0 0 1 2 0 30 0 0 3 3 00 10 8 4 0 30 0 0 5 3 00 12 6 6 0 30 0 0 7 3 00 14 4 8 0 30 0 0 9 3 00 16 2 10 0 30 0 0 11 3 00 18 0 12 0 30 0 0 13 3 00 19 8 14 0 30 0 0 15 3 00 21 6 16 0 30 0 0 17 3 00 23 4 18 0 3...

Page 70: ...f 216 92 Test profile Step Time min Speed km h Elevation Acceleration level 1 3 00 9 0 2 1 2 0 30 0 0 2 3 3 00 10 8 2 4 0 30 0 0 2 5 3 00 12 6 2 6 0 30 0 0 2 7 3 00 14 4 2 8 0 30 0 0 4 9 3 00 14 4 4 10 0 30 0 0 6 11 3 00 14 4 6 12 0 30 0 0 8 13 3 00 14 4 8 14 0 30 0 0 10 15 3 00 14 4 10 16 0 30 0 0 12 17 3 00 14 4 12 18 0 30 0 0 14 19 3 00 14 4 14 20 0 30 0 0 16 21 3 00 14 4 16 22 0 30 0 0 18 23 3...

Page 71: ... 8 5 5 1 9 1 00 14 4 0 1 10 1 00 10 1 0 3 11 2 30 10 1 3 3 12 2 20 7 0 9 1 13 0 10 9 2 5 1 14 1 00 15 1 0 1 15 1 00 10 8 0 3 16 2 30 10 8 3 3 17 2 20 7 7 9 1 18 0 10 9 9 5 1 19 1 00 15 8 0 1 20 1 00 11 5 0 3 21 2 30 11 5 3 3 22 2 20 8 4 9 1 23 0 10 10 6 5 1 24 1 00 16 5 0 1 25 1 00 12 2 0 3 26 2 30 12 2 3 3 27 2 20 9 1 9 1 28 0 10 11 3 5 1 29 1 00 17 2 0 1 30 1 00 12 9 0 3 31 2 30 12 9 3 3 32 2 20...

Page 72: ... able to control all functions of the running machine via host PC and also to serve for on line monitoring of all data During programming the displays indicate the values of the present program step and not the value of the total distance or total time programmed so far In case of entering a false value or in case you want to alternate a profile you can scroll the profile steps with the help of th...

Page 73: ...0 is flashing indicates program step St 1 05 Select speed 5 0 km h aor d 06 Confirm speed c indicates distance 0 is flashing 07 Confirm distance 0 this step is programmed after time c indicates time minutes 0 00 is flashing 08 Select time 5 minutes aor d 09 Confirm time c indicates time seconds 5 00 is flashing 10 Confirm time 00 seconds c indicates elevation Value 0 0 is flashing 11 Confirm eleva...

Page 74: ...25 Select time 4 minutes aor d 26 Confirm time c indicates time seconds 4 00 is flashing 27 Confirm time 00 seconds c indicates elevation 0 0 is flashing 28 Select elevation 5 aor d 29 Confirm elevation 5 c indicates acceleration level Acc 1 is flashing 30 Confirm acceleration level 1 c indicates speed 0 0 is flashing indicates program step St 4 31 Select speed 16 0 km h aor d 32 Confirm speed c i...

Page 75: ...lect time 5 minutes aor d 42 Confirm time c indicates time seconds 5 00 is flashing 43 Confirm time 00 seconds c indicates elevation 0 0 is flashing 44 Confirm elevation 0 c indicates acceleration level Acc 1 is flashing 45 Select acceleration level deceleration 2 aor d In this program step the speed will be reduced from 16 km h to 6 km h at an acceleration level 2 46 Confirm acceleration level c ...

Page 76: ...Keys Response Display 01 Select option mode OPxx for users Note The option mode for administrators is only accessible for authorised service technicians not for users a and simultaneously d and simultaneously b for at least 3 seconds indicates OP 01 flashing for option no 01 and indicating E rE SEt for error reset 02 Select option no 01 53 aor d indicates OP 01 OP 53 flashing and indicate a short ...

Page 77: ...ervice engineer The following intervals are set back by this operation E 01 Oil interval A OP 35 E 02 Service time interval A OP 37 E 02 Service distance interval A OP 38 OP02 Total distance covered km and indicates total distance covered in km indicates km OP03 Indication of total hours of operation stand by time including runtime of motor running belt h and show operation hours reports h OP04 In...

Page 78: ... or m min is blinking OP13 Unit for display of distance indicates unit for distance 0 km 1 miles 2 m m km or miles is blinking OP14 Unit for angle of elevation indicates Unit for elevation 0 per cent 1 degree or blinking OP15 Subject s body weight default value indicates 10 250 estimated weight weight blinking The personal body weight is necessary for a more correct calculation estimation of the p...

Page 79: ...ust not be used for new installations and configurations of medical devices and medical applications from 21 03 2010 h p cosmos para control and h p cosmos para graphics exclusively work with the h p cosmos coscom protocol indicates flashing number of the RS232 interface protocol and reports OFF RS232 not active no protocol interface deactivated 1 h p cosmos coscom v1 v2 v3 with Baudrate 9600 bps ...

Page 80: ...celeration level 4 The maximum acceleration and deceleration level is NOT valid for control and operation via V24 RS232 interface In this case the acceleration and deceleration level is set in option no 29 resp set in the corresponding command of the h p cosmos coscom protocol OP29 Standard acceleration and deceleration level for RS 232 interface The selected acceleration and deceleration level is...

Page 81: ...ART again or automatically after 2 minutes after having pressed STOP Default ON Display values will be continued added after pressing START again and will not be automatically deleted by pressing STOP Display values can only be deleted by pressing the STOP key twice time distance energy OP48 Countdown of program step OFF The time display counts up each program step ON The time display counts down ...

Page 82: ...t 65 kg 10 250 kg OP 16 Bodyweight request before manual automatic start OFF no request OFF ON OP 17 Unit of energy consumption JOUL kJoule JOUL kJoule CALO kcal OP 18 Maximum speed default in cardio mode 6 0 km h 0 0 max speed OP19 Setting of Polar W I N D system 0000 0000 xxxx xxxx OP 20 RS 232 interface protocol COM 1 1 h p cosmos coscom 1 20 OP 21 RS 232 interface protocol COM 2 1 h p cosmos c...

Page 83: ...lication The running machines designed for use in sports and fitness are not qualified for medical application N It is not allowed to utilize sports grade treadmills for medical application N It is not allowed to connect sports treadmills to medical devices such as ECG systems Operation The operating instructions for medical running machines can also be used for fitness chapter 5 entitled Operatio...

Page 84: ...t for you to fill out the form Therefore please fill out the form for registration immediately and send it back 7 1 Safety notes warnings precautions Five safety rules before start working N switch off disconnect from the mains N lock against reclosure secure against reconnection N check that lines and equipment are dead N ground and short circuit phases N cover partition or screen of adjacent lin...

Page 85: ... multiple plug socket is prohibited N An overload or voltage drop even temporary of more than 20 of the mains voltage might cause malfunctions and or defects and might totally switch off the running machine In the event of such high voltage drops or total power failure the running machine switches off and the running belt stops To start again switch on the running machine at the main ON switch and...

Page 86: ...ower plug when the unit is not used for long periods Keeping the connection may cause electric shock current leakage or fire due to the deterioration of insulation N If the device is to be stored unused in an unsupervised area for an extended period ensure that children do not have access N Do not put packing plastic bags plastic foils or other materials within reach of children as suffocation may...

Page 87: ...h p cosmos com created 30 01 2015 printed 30 01 2015 page 87 of 216 7 2 Electrical safety features Potential isolation transformer Potential isolation transformer saturn 450 300 Potential equalization pin connector Shielded motor cable Frequency inverter metal housing for shielding Opto coupler board cos13057 only with analogue inverter control Opto coupler board cos100601 only with digital RS485 ...

Page 88: ...pplier s components For this reason the worst case malfunction scenario has to be taken into account h p cosmos classifies the internal firmware and software components as class B due to a number of risk reducing design features and measurements Healthy people who practice sports are usually able to activate the emergency stop button of the treadmill and in this way switch off the treadmill quickl...

Page 89: ...npacking the machine and accessories read the instructions on the crate Make sure that the machine power connection cable or any optional equipment will not be damaged during unpacking Pay special attention to small parts so that you do not dispose of them or any instructions with the packaging Within Germany most devices are delivered and assembled by h p cosmos directly or by an authorised forwa...

Page 90: ...rarily fixed with a special transport fuse construction available at h p cosmos in order to avoid damage during transport Provided the right handrail is flat and temporarily fixed on the running surface the whole machine can be turned 90 to the right to an upright position This reduces the width of the machine so that it can be easily manoeuvred through narrow doors or windows A special transport ...

Page 91: ... p cosmos com created 30 01 2015 printed 30 01 2015 page 91 of 216 Transport with trolleys For easy transport put two normal trolleys under the device platform approximately 30 cm x 30 cm with 4 manoeuvrable wheels each to make the device mobile Trolley with 4 wheels for max 300 kg 660 lbs cos13016 Trolley with 4 wheels for max 500 kg 1100 lbs cos13672 Packaging of oversize running machines 200 30...

Page 92: ...tions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 92 of 216 Packaging of oversize running machines 450 300 Size of the treadmill box treadmill lengh 670cm width 450cm high 180cm transport by heavy load to the customer ...

Page 93: ... the help of a special basket and a mobile crane For the oversize machine saturn 450 300rs a heavy transport and crane is required Transports of heavy devices must only be carried out by authorised staff in compliance with the safety standards Otherwise there is serious danger for people and devices 7 4 3 Storage Store the devices at a temperature of 20 50 C All devices can be stored without power...

Page 94: ... transformers with high voltage connection as the electric and magnetic interference can falsify measurements or even render them impossible High voltage lines must be avoided nearby the device h p cosmos electrical devices with mains connections must neither be used in wet and humid areas e g swimming pools saunas etc nor in environmental chambers If not stated otherwise in the delivery informati...

Page 95: ...els of the elevation system may cause a print in the flooring Please use h p cosmos floor mats to protect sensitive flooring Floor protection mat 150 50 for running machines with deck size 150x50 cm cos14005 Floor protection mat 170 65 for running machines with deck size 170x65 cm cos14042 Floor protection mat 190 65 for running machines with deck size 190x65 cm cos14043 Other sizes available on r...

Page 96: ...ons within the building are not incumbent on the supplier h p cosmos N Read the nameplate of the machine before connecting the machine to the power outlet in order to determine whether the machine was designed for special voltage supply e g 110 V and 25 A The standard power supply of 230 V AC 50 60 Hz is sufficient for most of the running machines N The bigger running machines with a size of runni...

Page 97: ...se ask your electrical engineer or h p cosmos if you have further questions N Before installing the running machine please compare the specifications on the name plate concerning the mains voltage and the mains frequency with your local characteristics Connect only if identical N Check the main lead the voltage power supply outlet and ground wire protection contacts before plugging in Damaged lead...

Page 98: ...he corresponding plug pin next to the main switch on the front and then connect the mains plug N During electric safety measurements and test leakage currents etc the potential equalization cable is to be disconnected temporarily and re connected after the electric safety measurements and test 7 4 6 3 General notes and benefits of potential equalization The voltage of a conductor or body to earth ...

Page 99: ...doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 99 of 216 7 4 6 4 Plug assignment at the UserTerminal If the UserTerminal was disassembled during transport please connect the cables according to the following illustration Devices with MCU5 control unit since manufacturing date 02 2010 ...

Page 100: ...8hpc en 20150130_cos14310m5 v1 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 100 of 216 Devices with MCU5 control unit until manufacturing date 02 2010 ...

Page 101: ...8hpc en 20150130_cos14310m5 v1 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 101 of 216 Devices with MCU4 control unit until manufacturing date 12 2007 ...

Page 102: ...0_cos14310m5 v1 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 102 of 216 For devices with MCU4 control unit until manufacturing date 12 2007 and reverse belt rotation ...

Page 103: ...com created 30 01 2015 printed 30 01 2015 page 103 of 216 MCU5 control board and display board inside the UserTerminal Sticker for plug references at the back plate of the UserTerminal Sticker for plug references at the External UserTerminal control unit differs Ground connection at the UserTerminal Cable connection for ON OFF button and emergency stop in the UserTerminal GRD ground until manufact...

Page 104: ...eated 30 01 2015 printed 30 01 2015 page 104 of 216 7 4 7 Cable connections models 200 300 75 125 venus saturn cos100719 socket in UserTerminal cos100710 socket in UserTerminal for inverter drive and elevation motor cos100716 plug in running machine cos100721 plug in running machine 7 4 8 Cable connections model saturn 450 300 The above listed sockets and plugs have to be connected with the suppli...

Page 105: ... device Pay attention to safety standard of host devices Read manual Always use potential isolation Do not connect medical devices with sports devices DE Gefährliche Netzspannung Vor dem Öffnen des Gerätes Netzstecker ziehen Sicherheits Standard des Zubehörs beachten Anleitung lesen Immer Potentialtrennung verwenden Keine Medizinprodukte an Sportgeräte anschliessen FR Danger sous tension Avant ouv...

Page 106: ...o the respective IEC standards e g IEC 950 for data processing equipment and IEC 60601 1 for medical equipment Furthermore all configurations shall comply with the valid version of the system standard IEC 60601 1 1 IEC 60601 1 2 IEC 60601 1 4 and IEC 60601 1 6 Always use IEC 60601 1 approved potential isolation components when linking medical devices via interface Anyone who connects additional eq...

Page 107: ...norm requires the manufacturer to inform the responsible organization which means the applier hospital medical practice etc about the following risks in the technical description N The connection of a PEMS to a network or data sharing that includes other devices may lead to unknown risks for patients users or third parties N The applier hospital medical practice etc should determine analyse evalua...

Page 108: ...lt of the integration of a system and to ensure that remaining potential risks are identified during the operation of the system For the system integrator this means N he has to plan the integration of all ME devices or ME systems and all non medical devices according to the manufacturer s instructions N he has to perform the risk management for the integrated system N he has to forward any manufa...

Page 109: ...k N What predictable abuse can occur N Is the connection to the network or data network carried out in accordance with the intended use or purpose according to 3 para 10 of the Medical Devices Act N Could a false data flow occur from or to any connected PEMS N What should the medical data transferred through the network or data network achieve or what should be done with this data What happens if ...

Page 110: ... critical processes and applications for which a malfunction or interruption would lead to a critical situation for the patient as in an intensive care network are categorized as class C Such a network should not be connected to the general hospital network because such a connection might create uncontrollable risks The accessibility of such an isolated floating network has to be very high interru...

Page 111: ...ed explicitly in DIN EN 60601 1 3rd edition but the chapter on network and data network includes an approach to handling this development The standard requires that when using a network or data network with the aim of exchanging data between PEMS and PEMS or other IT devices such as servers and databases the manufacturer as well as the operator must have the necessary knowledge to build such netwo...

Page 112: ...ion is completed the instruction protocol has to be signed by the h p cosmos technician and all trained persons and returned together with the signed delivery note and the registration form to h p cosmos No Illustration Description 01 Compare the delivery note with the installed running machine Are all parts running machine accessories service box and running machine folder complete and without da...

Page 113: ...cosmos The potential equalization cable has to be connected AFTER these measurements 06 Lubricate the running machine with 30 ml special silicone oil which is part of the service box 07 Check the tension of the running belt no slip 08 Adjust the running belt properly before instructing the customer The allen key must be pulled out of the screw immediately after usage due to risk of injury Make sur...

Page 114: ...h a UserTerminal For these models an additional keyboard or an external UserTerminal for remote control via RS232 is available as an option see chapter entitled Accessories For maintenance and diagnosis we recommend the use of PC software h p cosmos para control Download www h p cosmos com en software para_control htm 8 1 Preventive maintenance The h p cosmos authorised service engineers are happy...

Page 115: ...ments and control steps consider also local requirements of your country and see detailed instruction and protocol form order no cos11690xx 8 3 1 Visual inspections N Carry out visual inspection for damages of device and complete accessories internal space engine compartment connecting lead and the right position of the tensile relief and plug ground wire connections etc N Carry out visual inspect...

Page 116: ...le or the connectors have been damaged In this case the cable should be replaced and the device repaired immediately 8 3 3 Isolation resistance RISO measurement N Resistance between bridged L N and Protective Earth Connection N Make sure that all isolations that are under stress of the mains voltage are considered All switches and contactors should be connected N The measurement is to be performed...

Page 117: ...for electric safety tester based on IEC 60601 1 For all measurements read operation manuals of the measurement devices carefully and verify the values and intervals with the local guidelines and laws An appropriate inspection record for all measurements and instructions is available at the manufacturers See also detailed instructions and special protocol cos11690xx Upon first installation at the c...

Page 118: ...ectricity circuit It should be pointed out that these are not regulations standards or safety checks specifically for h p cosmos running machines but rather apply to all electrical devices and building installations worldwide including all electrical devices with metal housing 8 4 Lubrication of the running belt running surface Do not touch the running belt while it is in motion This maintenance m...

Page 119: ...surface gliding on the deck surface of the running belt has a fabric construction which accumulates and distributes the oil until the next oiling is due Oiling amount Oil in the amount of 3 syringes each filled with 10 ml of silicone oil Special devices may have different oiling amounts see delivery papers In addition 1 syringe filled with air should be added in order to empty the oil line in the ...

Page 120: ...d of 5 km h so that the oil is distributed on the inside of the running belt Walk for 2 minutes at very low speed ca 2 km h while varying your position on the running belt This will allow the oil under the belt to distribute evenly 10 Check the position of the running belt after lubrication The belt should be in a centre position on the rear roller tail shaft If the belt needs to be adjusted pleas...

Page 121: ...ntity of oil absorbed by measuring the difference in weight However the user should realize that this test can only show approximate values and trends and is not meant to replace the direct inspection by an authorized h p cosmos technician at the prescribed time intervals For security reasons the running machine must be switched off and the power plug must be pulled for full separation from the ma...

Page 122: ...c oil pump Devices without a UserTerminal are equipped with an automatic acoustic signal bleep indicating the device needs to be lubricated At the standard setting the acoustic signal will occur after every 1000 km of operation Special devices may have different intervals After switching on the device using the main power switch the acoustic signal will be repeated three times 5 times LONG code fo...

Page 123: ...antity and quality of oil is especially important for running machines designed for cycling and wheelchair applications For those models a lubrication test tissue test according the instructions available from h p cosmos should be performed during the obligatory yearly maintenance The test should be sent back to h p cosmos for evaluation User cooperation and inspection obligation User cooperation ...

Page 124: ...nal control unit indicates that the oil tank is nearly empty Refill the oil tank when the control lamp illuminates At standard settings the word OIL will flash after 1000 km of operation If this occurs check the oil level in the oil tank and refill preterm if necessary Because the device does not have a sensor the oil message will not extinguish automatically After checking the oil level the messa...

Page 125: ...atic lubrication 0 25l bottle cos14007 Special oil for sliding plate with automatic lubrication 0 50l bottle cos13438 h p cosmos treadmill size 450 300 Special oil for sliding plate with manual lubrication 0 25l bottle cos10218 8 4 6 Refilling hydraulic oil tank N Switch off the treadmill N For refilling the hydraulic aggregate please open the marked lid N Please use hydraulic oil cos101422 N The ...

Page 126: ... while or if it has been adjusted wrongly In this case a backlash occurs between the driving shaft and the belt i e when weight is applied to the belt by stepping on it the belt slows down The belt tension should be checked as follows N Visually inspect the surface for cracks In case of a crack replace running belt immediately N Open the motor hood at the front Take care that nobody gets his hands...

Page 127: ...the tail shaft at the back of the device belt re entry zone Take care that long hair and loose clothing are not caught in the gap of the tail shaft N For safety reasons the adjustment procedure must be attended by a second person who can immediately press the emergency stop button in an emergency N Switch off the running machine and pull the mains plug before intervening N In case of any maintenan...

Page 128: ...g belt will remain in the adjusted position for a period of months if it has been adjusted correctly N During reverse belt rotation downhill run a re adjustment of the belt position centring is required N By evenly turning the left and right trimming screws to the right the running belt can be tightened if necessary See chapter running belt tension Note The running belt always has to be positioned...

Page 129: ...ollers and thereby the direction of the running belt The outmost light barriers channel 4 and 8 are connected to an opener If the running belt has moved too far to one side they are activated and the treadmill stops immediately In this case please contact the h p cosmos service hotline For cleaning the lightbarriers switch off the treadmill Reader 1 belt control Reader 2 Left side in belt directio...

Page 130: ...re entry zone Take care that long hair and loose clothing are not caught in the gap of the tail shaft N For safety reasons the adjustment procedure must be attended by a second person who can immediately press the emergency stop button in an emergency N Switch off the running machine and pull the mains plug before intervening N In case of any maintenance work on the treadmill ties have to be remov...

Page 131: ...uency during oscillation N Check the tone in the display target 1G adjust the tension of the drive belt if necessary N Check the amplitude of the needle target 440 460 adjust the tension of the drive belt if necessary N When the tuning device indicates the tone 1G and an amplitude of the needle of 440 460 the tension of the drive belt is OK Chart with tones to the corresponding frequency Target va...

Page 132: ...e cleansing disinfecting and before opening N Pay attention to the safety notes warnings and precautions of the running machine the accessories and the disinfectant The h p cosmos devices are neither sterile nor can they be sterilized Pay attention to the operating instructions of the respective devices and accessories Chemicals needed for application or cleansing should be stored in appropriate r...

Page 133: ...e shaft of the drive engine N devices manufactured since approx 2003 have hall effect sensors with metal toothed lock washer instead of light barriers N only for oversize running machines with a running surface of 200 75 cm up to 300 125 cm and built before 2002 the speed sensor is mounted in its own housing at the rear of the drive engine drive shaft thus it must not be cleaned The light barrier ...

Page 134: ...d unwanted activation the light barrier and the reflector glass should be cleaned carefully with a cloth moistened with alcohol at intervals of approximately one week depending on the air and dust conditions in the room Do not clean the light barriers at saturn 450 300 for belt control while the machine is powered on because of uncontrolled reactions of the belt control system 8 13 2 Adjustment of...

Page 135: ...l PC does not boot up it can be switched on manually Therefore the external control unit has to be opened in order to reach the power switch at the rear of the TouchPanel PC Opening of the device must only be performed by authorized technicians 9 2 Mechanical noise problems If knocking or rattling noises are heard during training please check if the machine has a firm stand and follow exactly the ...

Page 136: ...6 9 4 Fuses Device protection switch thermal protectet 16A ON OFF switch Expulsion fuse ON OFF button 2 pole C16A Expulsion fuse device fuse 3 pole C16A Secondary fuse isolation transformer 1 pole B16A Secondary fuse isolation transformer 3 pole C16A Expulsion fuse on devices with external switching cabinet Fuses in cabinet for power supply unit Fuses in UserTerminal saturn 450 300 Fuses of saturn...

Page 137: ...with up to several thousand volts If the user then touches a metal piece keys or display there may be an electrostatic discharge between the user and the device Such discharge may interfere with the operation of the device Generally electrostatic discharges are without harm for the user as well as for the device but can be quite unpleasant The main causes for electrostatic discharges are the choic...

Page 138: ...r sports or medical devices are defective e g because of a defective isolation inside the device the current flow may be fatal if bare metal parts for example screws are touched Disengage the device immediately unplug it and secure it against new start up Order an authorised electrician to repair the contactor circuit in the building or and on the device Read and follow the directives and informat...

Page 139: ...S232 interface N Loop back test For testing the RS232 of the running machine a special RS232 test plug including testing instruction is available from the manufacturer Connect the plug to the RS232 port and adjust OP20 10 N A blinking code will appear and the instructions will tell you if input and output of the RS232 are working well N PC software h p cosmos para control If you install the softwa...

Page 140: ...he device Error messages on devices with UserTerminal Initiation Running belt is not moving One of the mode LED is flashing manual profile cardio test Message Description Explanation Display Activity E01 Oil interval due E 01 OIL flashing HELP flashing Running machine is still functioning N Every 1000 km interval is adjustable lubrication service is due See chapter 8 4 Lubrication of the running b...

Page 141: ...CU is longer than the high acceleration command SPRINT function or high acceleration level N Trouble shooting see administrator OPTION 48 N Contact authorized service engineer N Check voltage supply Do not use extension cords Connect the machine directly to the socket in the wall N Check mechanical parts for any malfunction Several persons may have been standing on the machine for a moment N Remov...

Page 142: ...flashing Running machine is still working N At MCU5 control electronic as of manufacturing date 12 2007 this error shows a defect of the option memory FRAM In this case please contact h p cosmos and request a service technician N Optional setup restores to default factory setting and device type N By pressing the key c the device can still be used but it is likely that individual settings and prog...

Page 143: ...ncy Inverter Indication display UT Indication display FI Meaning of the error code related to the frequency inverter motor drive control function status Misubishi 00 No alarm 10 E OC1 Overcurrent shut off during acceleration 11 E OC2 Overcurrent shut off during constant speed 12 E OC3 Overcurrent shut off during deceleration or stop 20 E OV1 Regenerative overvoltage shut off during acceleration 21...

Page 144: ... zone 32 SCF3 Ground short circuit 33 OPF2 3 output phase loss 42 SLF2 SoMove communication 45 SLF3 HMI communication 51 InF9 Current measurement circuit 53 InFb Internal thermal sensor detected fault 54 tJF IGBT overheat 55 SCF4 IGBT short circuit 56 SCF5 Load short circuit 69 InFE Internal CPU 77 CFI2 Download invalid configuration 100 ULF Process underload fault 101 OLC Process overload 102 SPI...

Page 145: ...ror chip card drive not all models CARD ERR 2 Running machine is still functioning except chip card function Wrong PIN number N False chip card in the chip card drive or N Too many PIN retries wrong PIN configured Card Err 03 Error chip card drive not all models CARD ERR 3 Running machine is still functioning except chip card function N Wrong card type N no treadmill training session configured N ...

Page 146: ...y be reset in the administrator level In this case please contact customer service Error messages at devices without UserTerminal Devices without UserTerminal are equipped with an automatic error message code via acoustic signal beep indicating the device has a malfunction needs lubrication or regular service The appropriate beep code sounds in the moment the error occurs and afterwards when switc...

Page 147: ...te is too high N Check the oil temperature display adjusted to 80 C N Check the cooler of hydraulic aggregate is the fan running Run over limit switch max min The elevation system activated the hardware switch for max min position N Check the elevation sensor proper fixation Electrical ramp error The ramp isn t starting after command N Check whether the hydraulic unit is running N Check oil temper...

Page 148: ...pear again OP02 Selection of the OEM code By selecting the OEM code the functions and settings of the UserTerminal are adapted to the requirements of the listed OEM partner The following OEM codes can be selected 0 h p cosmos standard 1 JAEGER VIASYS CARDINAL HEALTH CARE FUSION 2 Ergo Fit 1997 98 3 Proxomed Kardiomed 4 KISTLER Gaitway 6 COSMED 9 SCHILLER Not all companies listed above are or were ...

Page 149: ...or safety purposes e g for beginners or in case the ceiling height of the room s ceiling is insufficient the max elevation of every treadmill model can be limited indicates maximum elevation adjustable range from 0 to maximum treadmill elevation OP08 Heart rate display interval indicates heart rate display interval in seconds adjustable range from 0 beat to beat 1 to 9 seconds Compare An original ...

Page 150: ...es running belt movement it activates the motor break for 10 seconds Remark The motor brake works only on running machines with integrated speed measurement The speed sensor must be switched on OP11 1 If the running machine is not equipped with a speed measuring system this can be retrofitted easily To activate the motor brake some parameters of the frequency inverter must be changed Ask the servi...

Page 151: ...ce interval for indication of required OIL service can be adjusted by means of OP 35 Standard 1000 km for treadmills 100 km for ladder ergometers OIL intervals are indicated with error code E01 OIL every 1000 km indicates oil interval in km standard 1 000 km is flashing adjustment range OFF 100 5 000 km Do not change the interval without prior communication with h p cosmos For the models h p cosmo...

Page 152: ...nged with the keys and has not yet been reached report current elevation angle recommended default setting OP40 Speed display as set point or true value 0 Display of speed as set point report target speed 1 Display of speed as true value resp flashing set point report if the set point was changed with the keys and has not yet been reached report current speed recommended default setting OP41 Eleva...

Page 153: ...tled List of running machines types N Make sure that the correct device type is set at administrator OP03 N Go to administrator OP 05 and set maximum possible speed N Set the device in the manual mode to the maximum possible speed by using key N Check the analogue voltage at the MCU FU control with a volt meter It should be 10 V Make a note of the voltage in the service report not possible with di...

Page 154: ...01 2 elevation control increment disc with light barrier 60 incr per motor rotation or hall effect sensor with 60 incr per motor rotation elevation element with a 5 mm spindle for 0 25 elevation h p cosmos saturn since 1994 3 elevation control increment disc with light barrier 60 incr per motor rotation or hall effect sensor with 60 incr per motor rotation elevation element with a 5 mm spindle for...

Page 155: ...rmed nO d EF S AVEd The key SW1 on the MCU 5 assembly board performs the same action as the loading of default values with the following exception if there are no default values saved the default values of the firmware will be taken over Following option values are not handled as default values and remain unchanged User OP 02 Total distance User OP 03 Total time of operation User OP 04 Total time ...

Page 156: ...orts h Difference to user option 04 the displayed value can also be changed OP95 MCU serial number Report and entering of 8 digit serial number with the following format YYMM NNNN YY 2 digit number of year of production lot MM month of production lot NNNN 4 digit serial number OP98 Setting of limitation of usage time Secret option OP99 Test profile 99 for endurance testing in the factory service W...

Page 157: ...n depending on the device type in OP03 0 max elevation OP 08 Heart rate display interval 0 beat to beat 0 or 1 9 seconds OP 09 Display test for all displays OP 10 Enter the serial number of the device 300 12 digit serial number OP 11 Activating and deactivating speed measurement depending on the device type in OP03 0 without speed measurement 1 with speed measurement OP 12 Setting of allowance for...

Page 158: ...ins 0 elevation maintains 1 elevation gets down to 0 OP 46 Display and setting of elevation for ladder ergometer 75 degree 0 180 degree OP 47 Reset of oil and service interval OP 48 Running belt speed at maximum frequency inverter speed depending on the device type OP 03 depending on the device type OP 03 1 00 99 99 km h OP 49 Type of FI control RS 485 analogue RS 485 OP 51 Error statistics of the...

Page 159: ...0 4165 30 0 0 119 0 200 xx 3 1 no limit 40 0 35 1 2617 40 0 3 120 0 180 3 2 50 0 1 2617 50 0 3 125 0 3 3 60 0 1 2617 60 0 3 100 0 3 4 80 0 1 2617 80 0 3 130 0 3 5 60 0 1 1700 60 0 3 100 0 3 6 80 0 1 1700 80 0 3 130 0 250 xx 300 xx 450 xx 4 1 no limit 40 0 25 1 2617 40 0 2 120 0 4 2 50 0 25 1 2617 50 0 2 125 0 4 3 60 0 25 1 2617 60 0 2 100 0 4 4 80 0 25 1 2617 80 0 2 130 0 4 5 40 0 27 1 2617 40 0 2...

Page 160: ...s max 1s min during heel strike Drive motor capacity 2200 VA Elevation motor capacity 480 VA Safety standards c IEC 60335 1 VDE 0700 IEC 60601 1 2 VDE 0701 ISO 20957 1 EN 957 6 n MDD directive 93 42 EEC 2007 47 EC IEC 60601 1 VDE 0750 VDE 0751 IEC 60601 1 2 ISO 20957 1 EN 957 6 Earth leakage current 1 0 mA 0 12 mA Touch leakage current n a Isolation transformer 1840 VA Degree of protection class I...

Page 161: ... lubrication system manual Number of stored programmes 6 training profiles 10 test profiles 8 user definable profiles RS232 Digital interface serial port USB converter optional COM1 standard COM2 optional COM3 optional standard for service COM4 optional Heart rate measurement option PC software not for medical applications h p cosmos para control included options h p cosmos para graphics h p cosmo...

Page 162: ...trike Drive motor capacity 3300 VA Elevation motor capacity not applicable 390 VA not applicable 390 VA 390 VA a not applicable 390 VA Safety standards c IEC EN 60335 1 VDE 0700 IEC EN 60601 1 2 VDE 0701 ISO 20957 1 EN 957 6 Machinery Directive 2006 42 EC n MDD directive 93 42 EEC 2007 47 EC IEC EN 60601 1 VDE 0750 VDE 0751 IEC EN 60601 1 2 ISO 20957 1 EN 957 6 Machinery Directive 2006 42 EC Earth...

Page 163: ...sation is to be expected Central lubrication system manual Number of stored programmes 6 training profiles 10 test profiles 8 user definable profiles RS232 Digital interface serial port USB converter optional COM1 standard COM2 optional COM3 optional standard for service COM4 optional COM1 COM2 standard COM3 COM4 n a Heart rate measurement POLAR 1 channel receiver ECG precisely POLAR W I N D PC so...

Page 164: ...ity 3300 VA Elevation motor capacity not applicable 500 VA not applicable 500 VA Safety standards c IEC EN 60335 1 VDE 0700 IEC EN 60601 1 2 VDE 0701 ISO 20957 1 EN 957 6 Machinery Directive 2006 42 EC n MDD directive 93 42 EEC 2007 47 EC IEC EN 60601 1 VDE 0750 VDE 0751 IEC EN 60601 1 2 ISO 20957 1 EN 957 6 Machinery Directive 2006 42 EC Earth leakage current 0 6 mA 0 2 mA Isolation transformer 2...

Page 165: ...f stored programmes 6 training profiles 10 test profiles 8 user definable profiles RS232 Digital interface serial port USB converter optional COM1 standard COM2 optional COM3 optional standard for service COM4 optional COM1 2 standard COM3 optional for service COM4 n a W I N D COM1 COM2 standard COM3 n a Touchscreen COM4 n a W I N D Heart rate measurement and heart rate dependent endurance control...

Page 166: ...te 3 x C 16 A at 400 V 3N Power input long time 2700 VA Power input momentary 5700 VA peak performances for milliseconds max 1s min during heel strike Drive motor capacity 4300 VA Elevation motor capacity 500 VA Safety standards n MDD directive 93 42 EEC 2007 47 EC IEC EN 60601 1 VDE 0750 VDE 0751 IEC EN 60601 1 2 ISO 20957 1 EN 957 6 Machinery Directive 2006 42 EC Earth leakage current 0 25 mA Is...

Page 167: ...mber of stored programmes 6 training profiles 10 test profiles 8 user definable profiles RS232 Digital interface serial port USB converter optional COM1 and COM2 standard COM3 optional standard for service COM4 n a used for POLAR W I N D COM1 and COM2 standard COM3 n a used for UserTerminal TouchScreen COM4 n a used for POLAR W I N D Heart rate measurement and heart rate dependent endurance contro...

Page 168: ...ry 15200 VA 66500 VA Drive motor capacity 11000 VA 30000 VA Elevation motor capacity 370 VA 18000 VA Safety standards n MDD directive 93 42 EEC 2007 47 EC IEC EN 60601 1 VDE 0750 IEC EN 60601 1 2 VDE 0751 ISO 20957 1 EN 957 6 Machinery Directive 2006 42 EC Earth leakage current 0 25 mA on request Isolation transformer 3 x 3000 VA 3 x 21000 VA Degree of protection class I M type B R IP 20 Overvolta...

Page 169: ...degree Degree 2 Normally only nonconductive pollution Temporary conductivity caused by condensation is to be expected Central lubrication system manual running machines with r have automatic central lubrication system Number of stored programmes 6 training profiles 10 test profiles 8 user defined profiles RS232 Digital interface serial port COM1 standard COM2 standard COM3 n a standard for service...

Page 170: ...ld be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines 2 kV for power supply lines 1 kV for input output lines Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a t...

Page 171: ...broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the running machine is used exceeds the applicable RF compliance level above the running machine should be observed to verify normal operation If ...

Page 172: ...a CE0476 via coscom v3 interface released for complete medical treadmill series ergospirometry system CORTEX CORTEX Biophysik GmbH Leipzig Germany ergo spirometry IIa CE0124 via coscom v3 interface released for complete medical treadmill series ergospirometry system Geratherm BlueCherry Geratherm GmbH Germany ergo spirometry IIa CE0118 via coscom v3 interface released for complete medical treadmil...

Page 173: ... a processor Pentium 1 8 GHz or higher is required 11 8Interface protocols h p cosmos has only approved the interface protocol coscom v3 for medical applications according to MDD directive 2007 47 EC EN 62304 EN 14971 mandatory since March 21 2010 All other interface protocols older versions coscom v2 coscom v1 coscomekg dll protocols of other treadmill manufacturers etc are no longer validated by...

Page 174: ...ciently in his work The arm supports with additional keyboard and additional stop button offers a wide range of applications e g exercising in gentle way for patients people with over weight orthopaedic problems or cardiovascular diseases h p cosmos arm support cos00098060044 for handrails with 10 angle h p cosmos arm support cos12013 for handrails with 0 angle for example long handrails with 2 pi...

Page 175: ... arm supports is only admitted for walking and not for running N While running the arm supports have to be folded vertically downwards on the outside 1 The arm supports must never project into the area of the running surface 2 12 2Additional keyboard and additional stop button for arm support and remote control To complement the arm supports and to increase safety an additional keyboard has been d...

Page 176: ...33 02 Handrail shortened at left side speed grip left cos14763 01 Handrail short cos11456 Handrail long rolled 2 pillars cos10166 Handrail detachable separable cos14192ral9007 Please quote the serial number of your treadmill for detailed article order number and price of the accessories since several different versions for different sizes may be available N The crossbar always needs to be position...

Page 177: ...accordingly e g in a patient file We have deliberately refrained from offering standard settings for particular heights because the adjustments can widely differ according to the posture and the individual situation of the testing person The torsional elements have a scaling at any joint which is afferent when turned in the direction of the running surface The adjustments of the height and width o...

Page 178: ...sure cylinders and noted accordingly e g in a patient file We have deliberately refrained from offering standard settings for particular heights because the adjustment can widely differ according to the posture and the individual situation of the testing person Safety precautions and operation N The max load is limited to 210 kg 460 lbs body weight N The positioning of the adjustable bar handrails...

Page 179: ...the lever 02 Push button Adjust the bar handrail to the desired height by pressing the push button 03 Move the bars horizontally to adjust the required width 04 Lock the locking lever by pushing it down 05 N At the rear end of the parallel bars two telescope tubes are mounted inside the tubes of the handrails They can serve the subject when mounting the running surface with a wheelchair N After en...

Page 180: ... be done under supervision N If the clamp handles are not locked sufficiently there is danger of squeezing due to unintentional movement of the therapist s seat or the foot support Open and overhanging clamp handles are a source of danger as well Therefore pay attention to always lock the fixation screws before using the seat and foot support N Especially during adjusting take care that neither yo...

Page 181: ...ining and supports the training itself N The unweighting can be levelled precisely to the needs of the patient and ensures an optimal training environment N High patient comfort for more motivation N The high class breathable and very comfortable unweighting vest has a physiological form and is easily adjustable This avoids dermal irritations and pressure marks to a large extent N The proper fitti...

Page 182: ...reedom of body motion They do not have straps undercutting during gait and have an optimal fitting due to the individually adjustable strap system Four different vests provide adaptation to all body sizes and forms A special vest for children is also available Additional straps for gait correction vest are included in the shipment The vest can be adjusted by the testing person h p cosmos airwalk v...

Page 183: ...hair on the treadmill and before using the wheelchair over ground or on the street Wheelchair stabilizer for deck length 200cm cos10227 250cm cos10226 300cm cos00096110031 Available for the h p cosmos running machine series with track size 200 75 cm and bigger with the mark r N Adjust the range limiters properly to limit the operation range on the deck and to ensure the wheelchair does not fall of...

Page 184: ... replaced every 6 months or earlier at the first sign of wear The length of the rope should be adjusted so that the subject cannot use the rear 30 of the running deck Put on the chest belt so that the h p cosmos logo is at the front In order to attach the carabiner put the vertical belt 1 below the shoulder belts 2 facing the body thereby unloading the seam at the joint 3 If the chest belt is put ...

Page 185: ...erse mode at the next start Change from forward to reverse rotation N Turn the key operated switch and start the machine with no one on the running belt and only when the belt is not moving N Observe the belt position for 2 minutes and correct the belt position with the belt adjustment screw see chapter 8 6 Adjustment centring of the running belt on page 127 It is normal for the belt to move to on...

Page 186: ...ll cord safety stop actuator does not prevent the subject from falling With reverse belt rotation downhill the patient must always be secured with the safety arch with chest belt system Furthermore a supervisor must be present The length of the pull cord should be adjusted so that the subject cannot use the rear 30 of the running deck It limits the range of motion of the subject to the 70 of the f...

Page 187: ... including external power supply 90 260 VAC to 12 VDC 5 A cos10002 02 UserTerminal TouchPanel remote control for table with foot for desktop with PC software h p cosmos para control connected via 5 meter cable to RS232 interface to the running machine including external power supply 90 260 VAC to 12 VDC 5 A without holder for handrail of the treadmill UserTerminal Display Keyboard MCU5 for h p cos...

Page 188: ... emergency stops If there is no socket for external emergency stops order the special rear panel terminal for additional keyboard cos10109 cos15294L10m incl 10 m cable no spiral cable please stipulate the serial number of the device cos15294L15m incl 15 m cable no spiral cable please stipulate the serial number of the device 12 19 Safety multiple socket The stable safety multiple socket can be use...

Page 189: ...rface In that case a USB to RS232 interface adapter cable is available at h p cosmos under order number cos12769 To control via USB interface the processor must be a Pentium 1 8 GHz or higher Interface cable RS232 for connection to PC for h p cosmos para graphics h p cosmos para control PC ECGs ergospirometry devices 5 9 4 8 3 7 2 6 1 PC COM1 oder COM2 CONNECTOR DB9 Buchse female Laufband treadmil...

Page 190: ...e N Pay attention to the advice regarding safety and warnings about possible interferences in the chapters on errors technical data EMC and installation in this manual N WARNING Heart rate monitoring systems may be inaccurate N Over exercising incorrect or forbidden use may result in serious injury or death N If the user feels faint sick dizzy or pain or in case of any other health problems stop e...

Page 191: ... system you should keep a gap of at least 100 cm between the devices to exclude interference Transmission problems N The setting POLAR logo upside down is recommended for approximately 10 15 of the subjects N The setting transmitter at the back of the subject is recommended for approximately 1 2 of the subjects N However always try the normal setting first Cleaning The chest belt is washable In th...

Page 192: ...red coloured STOP and QUICK STOP button on the UserTerminal ARE NO EMERGENCY BUTTONS which are disconnecting from the mains power supply They are no emergency control based on IEC 60601 1 2005 Most of the additional functions of para control require an interface connection via the sophisticated coscom v3 protocol which is integrated in all h p cosmos treadmill ergometers with MCU5 control electron...

Page 193: ... and endurance parameters online The treadmill can be controlled automatically The software is not designed for medical diagnostics or evaluation The treadmill data can support the doctor s decision Data can be converted by the software for data transfer to EXCEL and further evaluation e g by EXCEL or h p cosmos para analysis software Data transfer to an editor software for further data processing...

Page 194: ... well The clear design of the program allows an easy evaluation The comprehensive printout simplifies the interpretation and is the basis for best possible coaching The software is not designed for medical use diagnostics and evaluation The treadmill data can support the doctor s decision With this software it is possible to adapt the integrated test methods and to generate individual ones Test de...

Page 195: ...tabase of the most common gait and motion specific disorders with the possibility to transfer to the individual report N Integrated player for the analysis of several video clips N Video overlay of up to eight videos N Full screen mode N Various 2D analysis tools such as gridlock and grid N Flexible 2D painting tools such as rectangles tracking lines arrows and text fields N Tools and measurement ...

Page 196: ...ations N Remote control of all running machine functions during tests or exercise via interface from the PC N Demonstrations for large groups students audiences etc with the help of a big screen or digital projector N Control of multiple h p cosmos treadmill ergometers via one PC N Firmware update via upload to Flash Eprom of MCU 5 N Service maintenance work setup programming and display of error ...

Page 197: ... p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 197 of 216 h p cosmos para control 4 cos10071 V4 1 0 h p cosmos para control 5 cos10071 V5 Interface connection cable RS 232 5m cos00097010034 Interface connection cable RS 232 10m cos00097010035 USB to RS232 interface adapter cable cos12769 01 ...

Page 198: ...ree PC software h p cosmos para control is the result of extensive improvement and functional expansion of the well known version 4 1 In addition to the pure remote control function of h p cosmos treadmill ergometers via the PC the software now provides additional functions such as an option setting configuration tool with integrated storage function firmware update quick stop key cool down key pa...

Page 199: ...he appropriate tools eye protection dust mask etc Contact service h p cosmos com to receive the safety data sheet of the European Commission Directive 91 155 EEC h p cosmos running machine h p cosmos running machines consist of powder coated and galvanized metals from different producers and qualities stainless steel parts aluminium parts plastics rubber electronics with cables boards and condense...

Page 200: ..._h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 200 of 216 14 Appendix 1 Certificates Latest versions always http www h p cosmos com en company certificates htm 14 1Certificate of the TSA TÜV according to ISO 9001 ...

Page 201: ...50130_cos14310m5 v1 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 201 of 216 14 2Certificate of the TSA TÜV according to EN ISO 13485 ...

Page 202: ...8hpc en 20150130_cos14310m5 v1 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 202 of 216 14 3EC Certificate of the TSA TÜV ...

Page 203: ...ion of test 17 34 1054 seconds Heart rate at 500 m 158 bpm 1000 m 156 bpm 1500 m 160 bpm and 2000 m 155 bpm Average heart rate 158 156 160 155 4 157 bpm Fitness Index 420 50 x 0 2 1054 x 0 19338 157 x 0 56 105 1 882 x 2 6 420 10 203 8 87 9 29 7 x 2 6 430 368 9 61 An Fitness Index value less than 70 is below average average value 100 Conclusion In this case a subject needs some endurance training T...

Page 204: ...m are to be sent to h p cosmos Instruction on general operation 1 Hand over operating manual Inform about operating and service manual always keep one within reach of all users 2 Refer to safety notes warnings and precautions in this manual Place the print out of the safety instructions DIN A4 form in the delivery folder close to the treadmill Children are not allowed near the treadmill On the tre...

Page 205: ...ell as ON OFF button at the UserTerminal 5 Explain safety features that the delivered model is equipped with for example emergency stop quick stop safety harness etc Do not use the emergency stop button for normal switching OFF of the device 1 2 minutes warm up time delay before next power on for medical machines 6 Explain the display panels What does which display stand for 7 Explain manual opera...

Page 206: ...of test mode and the integrated test programs inform about interface and compatibility list Potential isolation for any interface linkage is obligatory in the field of medicine Only connect to coscom v3 compatible and authorized host equipment Do not connect to equipment or software which is not explicitly declared as compatible by h p cosmos and by the other manufacturer Inform about the h p cosm...

Page 207: ... 14 Correct adjustment of levelling sockets Check weight load on the levelling sockets by trying to lift up the treadmill frame 15 Proper adjustment of running belt with regulation screw and allen key For each change of running belt direction there are two adjustment steps a Turn key at the back of the UserTerminal b Adjust regulation screw according to previously defined value Pull out allen key ...

Page 208: ...y only be performed by authorised and trained personnel who are certified by h p cosmos 19 Inform the user about periodically recurring maintenance intervals and obligatory safety controls as well as the danger of disregarding these points Offer and recommend maintenance contract through authorized and trained personnel Advice and support documents 20 Hand over and explain delivery papers folder F...

Page 209: ...rd of intended and forbidden use safety notes or precautions as well as unauthorized maintenance or lack thereof and or regular safety checks may lead to injuries or even death and or can damage the device This will furthermore result in loss of liability and warranty Please fill out the instruction protocol and send it back to h p cosmos via Fax to 49 86 69 86 42 49 or via email to sales h p cosm...

Page 210: ...1_08hpc en 20150130_cos14310m5 v1 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 210 of 216 15 4Service report cos15531 ...

Page 211: ...8hpc en 20150130_cos14310m5 v1 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 211 of 216 15 5Control protocol cos11690en 02 ...

Page 212: ... 08hpc en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 212 of 216 15 6Wiring diagram load circuit quasar med pulsar Colour Code neither US nor Canadian standard ...

Page 213: ...c en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 213 of 216 15 7Wiring diagram control circuit quasar med pulsar page 1 Colour Code neither US nor Canadian standard ...

Page 214: ...c en_instructions_for_use_h p cosmos_treadmill doc 2015 h p cosmos sports medical gmbh author fh email h p cosmos com created 30 01 2015 printed 30 01 2015 page 214 of 216 15 8Wiring diagram control circuit quasar med pulsar page 2 Colour Code neither US nor Canadian standard ...

Page 215: ...ety information N Earth IEC 60417 5017 Warning safety precautions Pay attention to accompanying instructions ISO 7010 W001 Non sterile product 2012 Manufacturing date P Potential compensation IEC 60417 5021 Manufacturer Chassis ground R Applied part of type B IEC 60417 5840 Risk danger of entanglement Q Applied part of type BF IEC 60417 5333 Symbol for collection treatment recycling and disposal o...

Page 216: ...additionally Skype with webcam Service department phone 49 86 69 86 42 0 phone direct 49 86 69 86 42 25 mobile 49 171 720 69 88 fax 49 86 69 86 42 49 email service h p cosmos com skype h p cosmos com search select name Sales department phone 49 86 69 86 42 0 fax 49 86 69 86 42 49 email sales h p cosmos com youtube www youtube com hpcosmos twitter www twitter com hpcosmos facebook www facebook com ...

Reviews: