GOM-804 & GOM-805 User Manual
42
The
△
% compare function displays the deviation of the
measured value from the reference value as a percentage.
{ [(Measured Value-Reference)/Reference]%}.
Compare function
indicator
Deviation of the
measured value from
the reference value as
a percentage
Measured value
Reference, limits, compare
mode and beep mode
Pass/Fail judgment
The upper (HI) and low (LO) limits are set as a
percentage
from
the reference value. (Identical to the %
compare mode)
A measured value that falls within the upper and lower
limits is considered IN (pass), a value that falls below the
lower limits is considered LO, and a value that falls over
the upper limit is a HI.
Reference
value
Upper
limit
Lower
limit
IN
HI
LO
%
%
The % compare mode displays the measured value as a
percentage of the reference value [(Measured
Value/Reference Value)%].
The upper (HI) and low (LO) limits are set as a
percentage
from
the reference value. (Identical to the
△
% compare mode)
www.
GlobalTestSupply
.com
Find Quality Products Online at: