Note
The upper limit must be higher than the lower limit. Not
setting the upper limit higher than the lower limit is not
allowed. Likewise the lower limit cannot be set higher
than the upper limit.
5. Beep setting
Use the arrow keys to navigate to the Beep setting.
Press Enter to toggle the beep setting.
Beep setting
Move
Toggle
Enter
Beep Setting:
Off, Pass, Fail
Note
The Beep setting can also be set from the
System>Utility>Beep>Binning menu.
6. To start
binning
The binning function starts automatically if you are in
internal trigger mode.
If you are using the manual triggering mode, press the
Trigger
button or apply a pulse on the trigger pin of the
Handler interface to start binning.
See page 38 to set the triggering modes.
7. Display the
binning results
There are two different display modes to view results.
The Comp (Compare) display mode is the default display
mode. This mode will display the currently measured
value and displays which of the bins (if any) the
measured value is graded as.
Measurement
Grading results:
Green = IN
Red = OUT