8
disposable sleeve, flush into the marked recess on the upper side of the charging
base. Then activate the light and read the value displayed on the screen.
Measurement accuracy is in the range of +/–10%.
CAUTION:
If the tip of the top of main unit is placed obliquely to measure
the light intensity, there will be obvious deviation in the measured light intensity
value. The effective measurement of the integrated radiometer is the diameter of
the light source is 10 mm, if the measured light source diameter is not 10 mm
(As common 8mm), it will lead to the obvious difference in the measured light
intensity. At the same time, different manufacturers have different standards for
measuring the light intensity value. It is suggested that it should only be used
for measuring the original curing light. If it is not the original curing light, the
measured light intensity value may have obvious differences.
5.10 When operating, put the disposable sleeve on the main unit, clip the
notch of the light hood into the head of the top of main unit, and then clamp the
light hood down on the main unit. Aim the top of main unit at the correct position,
press the power button, a “di”sound will appear, the curing light shine the blue
light and start to work under the selected mode. Screen began to display the
countdown time. When the countdown back to 0 the work finished .Then the
screen will return to setting time.
5.11 While operation, press the power switch button can stop work at any
time.
5.12 At the end of a working cycle , the next working cycle can be started
immediately by short pressing the button. If the main unit gets hot obviously,
please turn off the device until the main unit becomes cool. Please don’t make it
continuously illuminate more than 10 times.
5.13 Low power detect circuit is fixed inside of the main unit, when low
power is detected, the battery level icon on main unit screen displays 0 grid
power and flickers, or "Low Battery!" is displayed. Please charge in time.
5.14 When charging, connect the adapter with the charging base. After the
screen front of the main unit and the screen front of the charging base in the
same direction, insert the main unit into the charging base and gently press
down to fasten the main unit and the charging base. At this time, the battery
Summary of Contents for O-Star CURING LIGHT
Page 1: ......
Page 23: ...ZMN SM 426 V1 0 20210609 ...