10
Prohibit aiming light at eyes directly.
6.10 Only the original adapter could be used, because other brand adapters
are likely to damage the circuit.
6.11 It is forbidden to use metal or other conductors to touch the charging
point of main unit, because it may burn the internal circuit or make the lithium
short circuit.
6.12 Charging the battery in the condition of cool and ventilated. Please
make sure of pressing out the buckle between the main unit and the pedestal,
otherwise the battery charging might be failed because of the poor contact.
6.13 Do not disassemble the Lithium battery, it will lead to the circuit short or
the electrolyte leakage.
6.14 Do not squeeze, shake and short the battery, do not store the battery
with metal material.
6.15 The instrument has electromagnetic interference. Do not use around the
electronic operation, at the same time have a strong electromagnetic interference
environment should be careful to use the instrument.
6.16 I t is forbidden to use when charging or operation.
6.17 This product should be used by trained, qualified dentists. And this
product is suitable for dental patients. Must be use in hospital or professional
medical site.
6.18 To avoid electromagnetic interference, the device should be installed at
the medical site which meet the requirement of EMC.
WARNING: The adapter should be connected to the socket which is easy for
operator to touch.
WARNING: over-heat scorching: The duty cycle of the equipment is 20 Sec
on/20 Sec off, if the curing light works for 40s continuously, the temperature of
the top of main unit may reach 56
℃
.
WARNING: parts of the ME equipment that are not serviced or maintained
while in use with the patient.
7. Cleaning, Disinfection and Sterilization
The cleaning, disinfection and sterilization of Point cure lens is as follow.
Unless otherwise stated, they will be hereinafter referred to as “products”.
Warnings
The use of strong detergent and disinfectant (alkaline pH>9 or acid pH <5)
will reduce the life span of products. And in such cases, the manufacturer takes
no responsibility.
This device shall not be exposed to high temperature above 138
℃
.
Processing limit
The products have been designed for a large number of sterilization cycles.
The materials used in manufacture were selected accordingly. However with
every renewed preparation for use, thermal and chemical stresses will result in