12
ISO15883. For the specific procedure, please refer to the automated disinfection
section in the next section “Disinfection”.
Notes
a) The cleaning agent does not have to be pure water. It can be distilled
water, deionized water or multi-enzyme. But please ensure that the selected
cleaning agent is compatible with the product.
b) In washing stage, the water temperature should not exceed 45 °C,
otherwise the protein will solidify and it would be difficult to remove.
c) After cleaning, the chemical residue should be less than 10mg / L.
7.4 Disinfection
Disinfection must be performed no later than 2 hours after the cleaning
phase.
Automated disinfection is preferred if conditions permit.
7.4.1 Automated disinfection-Washer-disinfector
·The washer-disinfector is proved to be valid by CE certification in accordance
with EN ISO 15883..
·Use high temperature disinfection function. The temperature does not
exceed 134
℃
, and the disinfection under the temperature cannot exceed 20
minutes.
·The disinfection cycle is in accordance with the disinfection cycle in EN ISO
15883.
Cleaning and disinfecting steps by using Washer-disinfector
1. Carefully place the product into the disinfection basket. Fixation of product
is needed only when the product is removable in the device. The products are
not allowed to contact each other.
2. Use a suitable rinsing adaptor, and connect the internal water lines to the
rinsing connection of the washer-disinfector.
3. Start the program.
4. After the program is finished, remove the product from the washer-
disinfector,
inspect (refer to section “Inspection and Maintenance”) and packaging (refer
to chapter “Packaging”). Dry the product repeatedly if necessary (refer to section
“Drying”).
Notes
a) Before use, you must carefully read the operating instructions provided by
the equipment manufacturer to familiarize yourself with the disinfection process
and precautions.
b) With this equipment, cleaning, disinfection and drying will be carried out
together.
c) Cleaning: (c1) The cleaning procedure should be suitable for the product
to be treated. The flushing period should be sufficient (5-10 minutes). Pre-wash
for 3minutes, wash for another 5 minutes, and rinse it for twice with each rinse