21
Recommended separation distances between
portable and mobile RF communications equipment and the device
The device are intended for use in electromagnetic environment in which
radiated RF disturbances is controlled. The customer or the user of the device
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the device are recommended below, according to the
maximum output power of the communications equipment.
Rated maximum
output power
of transmitter
W
Separation distance according to frequency of transmitter /m
150kHz to 80MHz
d=1.2×P1/2
80MHz to
800MHz
d=1.2×P1/2
800MHz to 2,5GHz
d=2.3×P1/2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) accordable to the transmitter
manufacturer.
NOTE I At 80 MHz and 800 MHz. the separation distance for the higher
frequency range applies. NOTE 2 These guidelines may not apply in all
situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
16. Statement
All rights of modifying the product are reserved to the manufacturer without
further notice. The pictures are only for reference. The final interpretation rights
belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD. The
industrial design, inner structure, etc, have claimed for several parents by
WOODPECKER, any copy or fake product must take legal responsibilities.