6
(Irrigation Mode) as the tip is vibrating.
7.6 The tip must be tightly screwed.
7.7 During operation, it is necessary to drip water for cooling. At the same
time, the water dripping instrument needs to be isolated from the earth.
7.8 The damage or wear of working tip will result in vibration intensity
decrease. The operator should replace the work tip according to the clinical
situation.
7.9 Don’t twist or polish work tip.
7.10 Don’t strike or scratch the handpiece.
7.11 When charging with the adapter, keep the contact with the adapter
housing for less than 1 min.
7.12 Please turn off the power switch after each treatment.
7.13 Woodpecker is specialized in producing medical instrument. We are
only responsible for the safety on the following conditions:
a) The maintenance, repair, and modification are made by the manufacturer
or the authorized dealers.
b) The charged components are original of “DTE” and operated according
to instruction manual.
7.14 The internal thread of Ultrasonic Endo activating file produced
by some other manufacturers maybe coarse, rusty, collapse, or use other
standard thread. It will be easy to be broken or cause slide teeth while
matching the external thread of handpiece, resulting in damage beyond repair
of the Ultrasonic Endo activating file. Please use “DTE” brand work tips.
8 Product contraindications
8.1 The hemophilia patient is forbidden to use this equipment.
8.2 The patients with heart pacemaker are forbidden to use this equipment.
8.3 The patients with heart pacemaker are forbidden to use this equipment.
8.4 Heart disease patients, pregnant women and children should be
cautious to use the equipment.
9 Cleaning, disinfection and sterilization
The cleaning, disinfection and sterilization of handpiece, tip, and Endo
wrench are as follow.
Unless otherwise stated, they will be hereinafter referred to as “products”.
Warnings
The use of strong detergent and disinfectant (alkaline pH>9 or acid pH
<5) will reduce the life span of products. And in such cases, the manufacturer
takes no responsibility.
Do not clean the handpiece with ultrasound cleaning machine.
The products may not be exposed to temperature above 138°C.
Processing limit
The products have been designed for a large number of sterilization cycles.