3.7 Working tip parameters:
Output tip vibration frequency: 45kHz±4kHz
3.8 Environment parameters:
Environment temperature: +5°C ~ +40°C
Relative humidity at 30% -75%
Atmospheric pressure:70kPa ~ 106kPa
3.9 Device safety classification:
Classified by operation mode: Short time operating device
Type of protection against electric shock: Class II equipment with internal
power supply
Degree of protection against electric shock: BF type applied part
Degree of protection against harmful ingress of water: Ordinary
equipment (IPX0)
Degree of safety application in the presence of a flammable anesthetic
mixture with air, oxygen, or nitrous oxide: Equipment cannot be used in
the presence of a flammable anesthetic mixture with air, oxygen, or nitrous
oxide.
4 Installation and removal method
4.1 Open the package, check whether the equipment is complete as per
the packing list, place the host on a solid plane, holding it directly facing the
operator, and then insert the handpiece. As shown below: