Tools: Nap-free soft cloth, tray
1. Soak the nap-free soft cloth with distilled water or deionized water, and
then wipe all the surfaces of the components such as the handpiece, host, tip,
etc. until the surface of the component is not stained.
2. Remove the tip from the handpiece, place it in a clean tray, and then
remove the handpiece from the host.
3. Wet the dry soft cloth with 75% alcohol, and then wipe all surfaces of
the host for 3 minutes.
4. Put the host back into the clean storage area.
Note:
a) The cleaning and disinfection must be performed within 10min before
use.
b) The disinfectant used must be used immediately, no foaming is allowed.
c) In addition to 75% alcohol, you can use non-residue disinfectants
such as Oxytech from Germany, but you must respect the concentration,
temperature and time specified by the disinfectant manufacturer.
10 Daily maintenance
10.1 This product does not contain self-repair spare parts. The machine
maintenance should be done by designated professionals or in the authorized
maintenance shop.
10.2 For this product, only handpiece, torch wrench, and tips can be
sterilized under high temperature of 134°C and high pressure of 0.22MPa.
For the other spare parts, they can be cleaned or sterilized by using water or
disinfectant to scrub their surface. Do not soak them in the solution. Do not
use volatile and diffluent solvent to clean as it will result in the color fading of
the marks on machine.
10.3 Please turn off the power switch, unplug the power plug when the
device is not in use. If not use for a long time, please make the device get
through to the power for five minutes once per month.