12
cabinet. The storage time should not exceed 7 days. If it is exceeded, it should
be reprocessed before use.
Notes:
a) The storage environment should be clean and must be disinfected
regularly;
b) Product storage must be batched and marked and recorded.
9.10 Transportation
1. Prevent excessive shock and vibration during transportation, and handle
with care;
2. It should not be mixed with dangerous goods during transportation.
3. Avoid exposure to sun or rain or snow during transportation.
The cleaning and disinfection of host and base are as follows.
1 Pre-Op processing
Before each use, the host must be cleaned and disinfected. The specific
steps are as follows:
Warning:
The host cannot be cleaned and disinfected with automatic
equipment. Manual cleaning and disinfection is required.
1
.
1 Manual cleaning steps:
1. Wet the soft cloth completely with distilled water or deionized water, and
then wipe all the surfaces of the host until the surface of the component is not
stained.
2. Wipe the surface of the host with a dry soft nap-free cloth.
3. Repeat the above steps at least 3 times.
Note:
a) Use distilled water or deionized water for cleaning at room temperature.
1.2 Manual disinfection steps:
1. Soak the dry soft cloth with 75% alcohol
2. Wipe all surfaces of host with a wet soft cloth for at least 3 minutes.
3. Wipe the surface of the host with a dry soft nap-free cloth.
Note:
a) The cleaning and disinfection must be performed within 10min before
use.
b) The disinfectant used must be used immediately, no foaming is allowed.
c) In addition to 75% alcohol, you can use non-residue disinfectants
such as Oxytech from Germany, but you must respect the concentration,
temperature and time specified by the disinfectant manufacturer.
2 Post-Op processing
After each use, clean and disinfect the host within 30 minutes. The specific
steps are as follows:
Summary of Contents for Endo 1
Page 1: ......
Page 24: ...ZMN SM 190 V1 8 20201103 ...