RF emissions
CISPR 11
Group 1
The model Endo 1 uses RF energy only for its
internal function.
Therefore, its RF emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11
Class B
The model Endo 1 is suitable for used in
domestic establishment and in establishment
directly connected to a low voltage power
supply network which supplies buildings used
for domestic purposes.
Harmonic
emissions
lEC 61000-3-2
Class A
Voltage
fluctuations
/ flicker emissions
lEC 61000-3-3
Compliance
17.3 Guidance & Declaration — electromagnetic immunity
Guidance & Declaration — electromagnetic immunity
The model Endo 1 is intended for use in the electromagnetic environment
specified below. The customer or the user of the model Endo 1 should assure
that It is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
lEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV,
±8kV, ±15 kV
air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power
supply lines
±1 kV for Input/
output lines
±2kV for power
supply lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
lEC 61000-4-5
±0.5 kV, ±1 kV
line to line
±0.5 kV, ±1 kV,
±2 kV line to
ground
±0.5 kV, ±1 kV
line to line
Mains power quality
should be that of a typical
commercial or hospital
environment.