11 Troubleshooting
Fault
Possible cause
Solution
No indication;
no action
The battery capacity of
Endo 1 Ultrasonic Endo
Activate Device is low.
Connect the power supply to
charge / replace the battery
The battery is damaged.
Replace the battery
Continuously using for
a l o n g t i m e , t h e r m a l
protection circuit action.
Stop using for few minutes,
and then it will function
normally.
Charging interface short
circuit results in lithium
b a t t e r y g e t t i n g i n t o
protection state.
Plug the device into the
charging base to charge, and
it will function normally.
Endo 1 Ultrasonic Endo
Activate Device is broken.
Contact with local dealers or
manufacturer.
Tips abnormally
working
Malfunction of host
Send to maintenance
department
Do not charge
after connecting
to power adapter
The power supply was not
well connected.
Pull out and connect again
T h e p o w e r s u p p l y i s
broken, or the specification
does not match.
Replace the battery
There are impurities on the
thimble of charging base.
Use ethyl alcohol to clean the
thimble on charging base.
The battery can
b e u s e d f o r a
shorter time after
charging
Battery capacity becomes
smaller
S e n d t o m a i n t e n a n c e
department
12 Storage and transportation
12.1 The equipment should be handled carefully and lightly. Be sure that
it is far from the vibration, and installed or kept in a cool, dry, and ventilated
place.
12.2 Do not store the machine together with articles that is poisonous,
combustible, caustic, or explosive.
12.3 This machine should be stored in a room where the relative humidity