5
4.3 When replacing the battery, please open the battery cover, first remove
the lithium battery, and then gently pull out the plug-in, put the new battery
connector plug correctly into the host, and then put new battery in. Put the
battery cover on the host mounting port, push up, and cover the battery.
4.4 For battery charging, plug the host into the charging base and press
tightly to make the host and base tightly match. While not charging, please
keep the charging base power off.
5 Operation
5.1 When using, first put the top of the tip into the proper position of the
root canal, click Mode switch button, and the machine will automatically adjust
to the best working state to work. The set irrigation time is 20s. After 20s, the
machine automatically enters the sleep mode. After changing the irrigating
solution, press the switch again to enter the irrigation mode again. Under
normal circumstances, during a complete root canal irrigation, operators needs
to replace the irrigation solution for 7-8 times.
5.2 The tips, Endo wrench and handpiece are removable and can be
autoclaved under high temperature of 134°C and high pressure of 0.22MPa.
5.3 Low voltage indication: If the light of mode switching button turns dark
or the power is not enough, it means that the battery capacity is low and it
needs to charge immediately.
6 Charging instructions
6.1 Use the corresponding charging base: Connect the power adapter
with charging base, and plug the host into the charging base. While charging,
the mode switching button light would flash. If the light turns to be normally on,
it means that battery charging is over. Generally a charging process lasts for 1.5
hours. And under full charged condition, the machine can continuously work
for 1 hour.
6.2 The battery of this product is of no memory. It can be used while
charging.
6.3 Before first use of this machine, please charge for at least 4 hours.
7 Safety precautions
7.1 Keep the machine clean before and after operation.
7.2 Operators should be equipped with adequate protections such as
goggles, masks, etc. to prevent cross-infection while operating.
7.3 The use of the product must meet the requirements of the relevant
operation normalization and relevant regulations of medical department, and
the use is limited to the trained doctors or technicians.
7.4 The tip and endodontic wrench must be sterilized before each
treatment.
7.5 Don’t install or remove the handpiece or tip under working mode