16
Humidity limit for storage: 10% ~ 93%
Atmospheric pressure for storage: 70kPa—106kPa
Temperature limit for storage: -20°C ~ +55°C
Authorised Representative in the EUROPEAN COMMUNITY
17 Electromagnetic compatibility
Attention:
1) Without Woodpecker agreement and authorization, private modification
of device may result in the electromagnetic compatibility problem of that
device or other devices.
2) The design and test of Endo 1 Ultrasonic Endo Activate Device
complies with the related operation regulations of electromagnetic
compatibility.
Note: In the case of an electrical fast transient burst test, there may be
a situation in which the power cannot be adjusted due to the interference of
the touch key. This does not affect the output of the whole power and is self-
recoverable after the test is completed. According to the experienced clinician
and those professionals who are capable of using specific device or system,
this risk is acceptable.
17.1 Electromagnetic compatibility key components
The electromagnetic compatibility key components of this product
consist of power cord, circuit board, IC chip. The use or replacement of non-
supporting accessories, cables, transducers, etc. will result in a significant
reduction in electromagnetic compatibility emission and immunity performance.
Please do not privately replace the machine parts.
17.2 Guidance and manufacturer’s declaration - electromagnetic emissions
Guidance and manufacturer’s declaration - electromagnetic emissions
The model Endo 1 is intended for use in the electromagnetic environment
specified below. The customer or the user of the model Endo 1 should assure
that it is used in such an environment.
Emmission test
Compliance Electromagnetic environment – guidance
Summary of Contents for Endo 1