Disassembling steps
9.3 Cleaning
The cleaning should be performed no later than 24 hours after the
operation.
The cleaning can be divided into automated cleaning and manual cleaning.
Automated cleaning is preferred if conditions permit.
9.3.1 Automated cleaning
• The cleaner is proved to be valid by CE certification in accordance with
EN ISO 15883.
• There should be a flushing connector connected to the inner cavity of the
product.
• The cleaning procedure is suitable for the product, and the irrigating
period is sufficient.
• Do not clean the handpiece with ultrasound.
It is recommended to use a washer-disinfector in accordance with EN ISO
15883. For the specific procedure, please refer to the automated disinfection
section in the next section "Disinfection".
Notes:
a) The cleaning agent does not have to be pure water. It can be distilled
water, deionized water or multi-enzyme. But please ensure that the selected
cleaning agent is compatible with the product.
b) In washing stage, the water temperature should not exceed 45 °C,
otherwise the protein will solidify and it would be difficult to remove.
c) After cleaning, the chemical residue should be less than 10mg / L.