27
5.9.9 Indicating age of ClO
2
in the reactor and in reservoir
tank
1. MAIN MENU > SERVICE > [OK].
2. ClO
2
PRODUCTION > [OK].
3. AGE OF ClO
2
> [OK].
The factory setting for both is 00:00 (minutes and seconds).
4. Press [Esc].
5.9.10 Testing the display
1. MAIN MENU > SERVICE > [OK].
2. TEST DISPLAY > [OK].
The test function is started. The display goes completely dark so
that each pixel can be checked. In addition, all LEDs are switched
on. They light up orange, and the red alarm LED flashes. After
about 5 seconds, the display returns to the SERVICE - TEST
DISPLAY submenu.
PRODUCTION ClO2
CYCLES
CHEMICALS
MAINTENANCE
AGE OF ClO
2
AGE OF ClO2
REACTOR 03:16
STOCK 00:00