Page 56
TTFM 6.1 Transit Time Flow Meter
Possible Causes:
Corrective Action:
NO ECHO INDICATION Icon: No Echo (cont.)
Empty pipe or partially filled
Pipe must be fluid filled and acoustically transparent in order to
obtain echoes.
Coupling compound washed out, or
sensor loose on pipe.
Remount sensor
Use Super Lube® Silicone Compound
SENSOR CONNECTIONS
OPEN/SHORT SENSOR ICON
No sensors attached
Short in transducer, or in triax transducer cable. Follow Sensor
Connections steps
Sensor Connections
Check sensor connections at TTFM and at sensor junction box.
Note: Refer to Sensor Cable Resistance Test to test final
connections.
METER READING HIGHER THAN EXPECTED?
Calibration Error
Review calibration menu. Pipe dimensions and fluid
selection/fluid velocity.
Higher flow rate than expected
Investigate pump/valves. Compare velocity with alternate
instrument.
Erratic measurement (set damping to
0% to check) due to electrical noise
or poor signal quality.
Ensure all Flowmeter wiring is in METAL conduit and sensor
shield is properly grounded.
Ensure correct power input Ground connection (<1 ohm
resistance).
Ensure 4-20mA Shield connected to Instrument Ground stud.
Contact Greyline for further assistance.
High viscosity fluid
Laminar flow profile due to high viscosity fluid requires an
adjustment to Cal Const.
Summary of Contents for TTFM 6.1
Page 2: ...Note This page has been left blank intentionally ...
Page 5: ...Page 5 TTFM 6 1 Transit Time Flow Meter 100 240 VAC Meter CONNECTIONS ...
Page 27: ...Page 27 TTFM 6 1 Transit Time Flow Meter TYPICAL SENSOR INSTALLATION ...
Page 53: ...Page 53 TTFM 6 1 Transit Time Flow Meter SENSOR MOUNTING COUPLING RECOMMENDATIONS ...