Page 36
TTFM 6.1 Transit Time Flow Meter
2. Install one of the stainless steel mounting brackets on the pipe. This will be the stationary bracket
not being rotated to the opposite side of the pipe. For a horizontal pipe, position the bracket at 3
or 9 o’clock. Hold the level up to the top of the bracket to ensure the angle is correct. For vertical
pipes, the bracket can be at any orientation about the pipe. Tip: Use a 5/16” nut driver to tighten
the hose clamp. Install the second bracket at approximately the distance specified by the Sens
Space value in the Setup menu.
3. Use alignment bar to ensure that the brackets are parallel. Completion of steps 2 and 3 is shown
below.