40
Operating manual for the TG uni 1, TG euro 1 and TG euro 1 med appliance tester
Conducting tests in accordance with DIN VDE 0701-0702, Example 3 – Devices with SELV / PELV (PC III)
Fig. 71
Fig. 72
Fig. 70
f
f
Test settings
Fig. 71 shows the Main menu for test object protective measures.
After selecting the protective measures by pressing the [SELV/PELV (Pc III)]
function button you reach the „Settings for measuring“ menu (Fig. 72).
With the function button [hand/auto] you determine whether the test
procedure should be carried out manually (by hand) or automatically.
The headline (upper right) shows the setting:
Manually by hand (m), automatic (a)
Manual test procedure means that you have to confirm every test
stage by pressing the [OK] button to get to the next test stage.
Automatic test procedure means that a test stage automatically
changes to the next after 5 seconds – except for confirmation of
connection to the mains voltage (see Point 6.6, basic settings of
the appliance tester).
The [cancel] button takes you back to the Main menu.
To move to the next stage press the [next] button (also see Diagram 1 in
Point 9.8).