VEGA

 it’s a device working 230V ~ / 50 Hz network electricity, to be used for the nasal aspiration, oral aspiration, 

tracheal aspiration of the body liquids (mucus, catarrh or blood) in the adult or in the children. Easily portable 
equipment designed for not continuous use 30 min. ON / 30 min. OFF. 
Thanks to this characteristics and to the rating that it has, this product is particularly suitable for hospital use, on the 
tracheotomized patients, minor surgical applications and post-operative therapy at home.  
Made of highly heat resistant, electrically insulated plastic material in conformity with the latest European  
safety standard, the product is supplied with a complete polycarbonate autoclavable jar with overflow valve and it is 
equipped with aspiration regulator and vacuum indicator located on the front panel. 

 

GENERAL WARNINGS 

 

 

 

READ INSTRUCTION MANUAL CAREFULLY BEFORE USE 

 

 

ONLY HIGHLY QUALIFIED STAFF USE RESERVED 

 

 

THE INSTRUMENT MUST NOT DISASSEMBLED 

FOR TECHNICAL SERVICE ALWAYS CONTACT GIMA 

 

IMPORTANT SAFETY RULES 

 

1. 

When opening packing verifies the device state: always check the presence of damages on the plastic 
surface, this can expose under tension internal parts. Also check the main cable integrity.                                                              

In case of damage do not connect the plug to the socket. Check integrity before every use.

 

2. 

Before connecting the instrument, always check that the electrical requirements on manufacturer label and 
the plug used comply with the electrical network used. 

3. 

Keep special attention to: 

 

Use only original accessories; 

 

The device can be used only with the bacteriological filter; 

 

Place instrument on stable and flat surfaces and to avoid the blockage of the air inlets on sides; 

 

Do not use in presence of air inflammable anaesthetic, oxygen or nitrous oxide; 

 

Do not touch the instrument with we hands, and avoid contact with liquids; 

 

Keep off the reach of children or not capable people without supervision; 

 

If not used do not leave the instrument connected to network; 

 

Do not pull cable to remove plug from socket; disconnect hold plug with fingers; 

 

Use and keep the instrument in a safe environment, protected from bad weather condition and keep 
off excessive heating; 

4. 

The device must be used for the use it was designed.                                                                                                         

A different use must be considered improper and dangerous. Manufacture cannot be considered responsible 
for damages due to improper use or connection to an electrical system not complying with current 
regulations. 

Do not use for anything else but the use defined by manufacturer;

 

5. 

Instrument and accessories discharging must be done following current law regulations in every country of 
use. 

6. 

For repairs, exclusively contact technical service and request the use of original spare parts.                                                             
Failure to comply with the above can jeopardise the safety of the device;

 

7. 

Particular precautions must be made concerning electromagnetic compatibility. The medical device must be 
installed and used according to information supplied with the accompanying documents; 

 

 

IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC 

DIRECTIVE 2002/96/EC: 

At the end of its working life, the product must not be disposed of as urban waste. 

It must be taken a special local authority differentiated waste collection or to a dealer providing this service. 

Disposing of a household appliance separately avoids possible negative consequences for the environment and 

health deriving from inappropriate disposal and enables the constituent materials to be recovered to obtain 

significant saving in energy and resources. As a reminder of the need to dispose of separately the product is 

marked with a crossed-out wheeled dustbin. 

 

 

8

Summary of Contents for SUPER VEGA BATTERY

Page 1: ...AL MODE D EMPLOI HANDBUCH MANUAL DE ISTRUCCIONES M28210 M Rev 3 10 07 GIMA spa Via Monza 102 20060 Gessate MI Italia ITALIA Tel 02 953854 1 Fax 02 95381167 E mail gima gimaitaly com www gimaitaly com...

Page 2: ...Utilizzare solo accessori originali Utilizzare sempre l apparecchio con il filtro antibatterico Posizionare l apparecchio su superfici piane stabili e libere da oggetti che possono ostruirne le prese...

Page 3: ...mbiente 5 35 C Percentuale umidit ambiente 30 75 RH Altitudine 0 2000m s l m Condizioni di conservazioni Temperatura ambiente 40 70 C Percentuale umidit ambiente 10 100 RH SIMBOLOGIA Apparecchio con C...

Page 4: ...nte e o l utente dell Aspiratore chirurgico VEGA devono assicurarsi che l apparecchio sia utilizzato in tale ambiente Prova di Immunit Conformit Guida all ambiente elettromagnetico Scariche elettrosta...

Page 5: ...aver tolto la gabbietta il galleggiante con guarnizione rossa rimosso da tappo vaso 4 Per disinfettare gli accessori utilizzare alcool denaturato o soluzione a base di ipoclorito 5 A pulizia ultimata...

Page 6: ...po del vaso avvitato male Svitare e riavvitare correttamente il tappo 2 Mancata aspirazione Guarnizione del tappo non in sede Svitare il tappo e sistemare la guarnizione in sede 3 Mancata chiusura del...

Page 7: ...alla formazione di schiuma all interno del vaso di raccolta svitare il coperchio dal vaso e riempire quest ultimo per 1 3 di acqua per facilitare le operazioni di pulizia e velocizzare la depressione...

Page 8: ...naesthetic oxygen or nitrous oxide Do not touch the instrument with we hands and avoid contact with liquids Keep off the reach of children or not capable people without supervision If not used do not...

Page 9: ...sion 350 x 210 x 180 mm Functioning to 35 C and 110 operating voltage 30 min ON 30 min OFF Working condition Room temperature 5 35 C Room humidity percentage 30 75 RH Altitude 0 2000m s l m Conservati...

Page 10: ...of the surgical aspirator VEGA should assure that it s used in such an environment Immunity Test Compliance Electromagnetic environment guidance Electrostatic discharge ESD IEC EN 61000 4 2 6kV on con...

Page 11: ...use either denatured alcohol or hypochlorate based solutions easily purchasable from chemist shops 5 Once cleaning has been completed insert the float in the cage with the seal facing towards the cage...

Page 12: ...w the cap and insert the seal properly in its seat 3 The float doesn t close If the cap has been washed ensure that the float is not partially detached Insert the float into it s place 4 The float doe...

Page 13: ...must be used on a plan of horizontal operation Connect the long silicon tube 4 to the other jar s lid connector 5 Connect the other end of the long silicon tube 4 to the probe plastic connector 6 then...

Page 14: ...el il doit tre branch 3 Faire tout particuli rement attention Utiliser uniquement des accessoires d origine Utiliser toujours l appareil avec le filtre antibact rien Placer l appareil sur des surfaces...

Page 15: ...210 x 180 mm Fonctionnement 35 C 110 alimentation lectrique 30 ON 30 OFF Conditions de service Temp rature ambiante 5 35 C Pourcentage humidit ambiante 30 75 RH Altitude 0 2000 m s m Conditions de con...

Page 16: ...e l aspirateur VEGA doivent s assurer que l appareil est utilis dans un tel environnement Test d Emission Niveau de test Guide l environnement lectromagn tique D scharges lectrostatiques ESD IEC EN 61...

Page 17: ...e enlev du couvercle du bocal 4 Pour d sinfecter les accessoires utiliser de l alcool d natur ou une solution base d hypochlorite 5 Apr s nettoyage monter le flotteur dans la cage en tenant le joint t...

Page 18: ...t le couvercle 2 L appareil n aspire pas Joint du couvercle pas en place D visser le couvercle et remettre le joint en place 3 Le flotteur bloqu Pr sence des incrustations sur le flotteur D visser le...

Page 19: ...rcle du bocal et remplir ce dernier de 1 3 d eau pour faciliter les op rations de nettoyage et rendre plus rapide la d pression pendant le fonctionnement puis revisser le couvercle sur le bocal Pour l...

Page 20: ...olgendes achten Nur Originalzubeh r verwenden Das Ger t darf nur mit dem Bakterienfilter verwendet werden Das Ger t auf einer ebenen und stabilen Fl che aufstellen auf der sich keine die Luftschlitze...

Page 21: ...flasche 16 l min Gewicht 2 5 kg Abmessungen 350 x 210 x 180 mm Betriebsar 35 C 110 Netzspannung 30 min ON 30 min OFF Betriebsbedingungen Raumtemperatur 5 35 C Raumfeuchtigkeit 30 75 RF H he 0 2000m N...

Page 22: ...es Absauger VEGA muss sich vergewissern dass das Ger t unter diesen Bedingungen benutz wird Emissionstests Pr fniveau F her elektromagnetische Umgebung Elektrosatische Entladungen ESD IEC EN 61000 4 2...

Page 23: ...ntfernen konischer Teil mit 3 ffnungen 3 Nach der Entfernung des K figs wird der Schwimmer mit der roten Dichtung vom Beh lterstopfen entfernt 4 Zum Desinfizieren der Zubeh rteile verg llten Alkohol o...

Page 24: ...bt Beh lterstopfen abschrauben und wieder richtig aufschrauben 2 Keine Absaugung Stopfendichtung sitzt falsch Stopfen abschrauben und Dichtung richtig anbringen 3 Das Schwimmerventil schlie t nicht Da...

Page 25: ...us der Netzsteckdose ziehen Den Deckel des Beh lters abdrehen und den Beh lter zu 1 3 mit Wasser f llen um das Reinigen und den Vakuumaufbau w hrend des Betriebs zu beschleunigen Den Deckel dann auf d...

Page 26: ...ctrica 3 Prestar particular atenci n a Utilizar s lo accesorios originales El aparado solo se puede utilizar con filtros bacteriologicos Posicionar el aparato sobre superficies planas estables y libr...

Page 27: ...mm Funcionamiento 35 C y 110 Alimentacion el ctrica 30 min ON 30 min OFF Condiciones de funcionamiento Temperatura ambiente 5 35 C Porcentaje de humedad ambiente 30 75 RH Altitud 0 2000m s n m Condici...

Page 28: ...El cliente y o el usuario del aspirador VEGA deben asegurarse que el aparato se utilice en dicho tipo de ambiente Test de Inmunidad Conformidad Gu a al ambiente electromagn tico Descargas electroest t...

Page 29: ...ado la jaula se quitar el flotante con junta roja del tap n vaso 4 Para desinfectar los accesorios utilizar alcohol desnaturalizado o soluci n a base de hipoclorito 5 Finalizada la limpieza introducir...

Page 30: ...Desenrosque la tapa y vuelvala a enroscar correctamente 2 Falta aspiraci n Junta del tap n fuera del alojamiento Desenroscar el tap n y acomodar la junta en su alojamiento 3 Falta cierre del flotante...

Page 31: ...or 7 en la posici n I para encender Para apagar poner el interruptor en la posici n 0 y extraer el enchufe de la toma de alimentaci n Desenroscar la tapa del vaso y llenarlo por 1 3 con agua para faci...

Page 32: ...DIREZIONE FLUSSO FLOW DIRECTION Montaggio Filtro Filter Assembling Mod TOBI HOSPITAL TOBI HOSPI PLUS cod 28237 TOBI CLINIC SUPER TOBI CARRELLATO Cod 28239 DIREZIONE FLUSSO FLOW DIRECTION Connessione V...

Reviews: