10
Guidance and manufacturer’s declaration – electromagnetic Emissions
(
as request by regulation
EN 60601-1-2:2001)
The surgical aspirator VEGA is intended for use in the electromagnetic environment specified below.
The customer or the user of the surgical aspirator VEGA should assure that it’s used in such an environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
Irradiated / Conducted
emissions
CISPR11
Group 1
The surgical aspirator VEGA only
used RF energy only for its internal
functioning. Therefore, its RF emissions
are very low and are not cause
interference in proximity of any
Electronic appliances.
Irradiated / Conducted
emissions CISPR11
Class [B]
Harmonic emissions
IEC/EN 61000-3-2
Class [A]
Voltage fluctuations / flicker
emissions
IEC/EN 61000-3-3
Complies
The surgical aspirator VEGA can
be used in all environments, including
domestic and those connected directly to
the public mains distribution that supplies
power to environments used for domestic
scopes.
Guidance and manufacturer’s declaration – electromagnetic Immunity
(
as request by regulation
EN 60601-1-2:2001)
The surgical aspirator VEGA is intended for use in the electromagnetic environment specified below.
The customer or the user of the surgical aspirator VEGA should assure that it’s used in such an environment.
Immunity Test
Compliance
Electromagnetic environment -
guidance
Electrostatic discharge (ESD)
IEC/EN 61000-4-2
± 6kV on contact
± 8kV in air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient / burst
IEC/EN 61000-4-4
± 2kV power supply
Mains power quality should be that of a
typical commercial environment or
hospital.
Surge
IEC/EN 61000-4-5
± 1kV differential mode
Mains power quality should be that of a
typical commercial environment or
hospital.
Loss of voltage, brief voltage
interruptions and variations
IEC/EN 61000-4-11
5%U
T
for 0.5 cycle
40%U
T
for 05 cycle
70%U
T
for 25 cycle
<5%U
T
for 5 sec
Mains power quality should be that of a
typical commercial environment or
hospital. If the user of the surgical
aspirator VEGA request
that the appliance operates continuosly,
the use of a continuity unit is
recommended.
Magnetic field
IEC/EN 61000-4-8
3A/m
The power frequency magnetic field
should be measured in the intended
installation location to assure that it’s
sufficiently low.
Conducted Immunity
IEC/EN 61000-4-6
3Vrms 150kHz to 80MHz
(for appliances that aren’t
life - supporting)
-
Irradiated Conducted
IEC/EN 61000-4-3
3V/m 80MHz to 2.5 GHz
(for appliances that aren’t
life – equipment)
-
Nota U
T
is the value of the power supply voltage