9
TECHNICAL CHARACTERISTICS
Model VEGA
Typology (MDD 93/42/EEC)
Medical Device Class II a
Classification UNI EN ISO 10079-1
HIGH VACUUM / LOW FLOW
Main Voltage
230 V ~ / 50 Hz
110 V ~ / 60 Hz
(no CE 0123)
Power consuption
184 VA
105 VA
Fuse
F 1 x 1.6A 250 V
F 1 x 4 A 250 V
Maximum suction aspiration (without jar)
-75kPa (- 0.75 bar)
Adjustable from -75kPa (-0.75bar) to -10kPa (-0.10 bar)
Maximum flow (without jar)
16 l/min
Weight 2.5
Kg
Dimension
350 x 210 x 180 mm
Functioning
(to 35°C and 110 % operating voltage)
30 min ON / 30 min OFF
Working condition
Room temperature: 5 ÷ 35 °C
Room humidity percentage: 30 ÷ 75 % RH
Altitude: 0 ÷ 2000m s.l.m
Conservation condition
Room temperature: - 40 ÷70 °C
Room humidity percentage: 10 ÷ 100% RH
SYMBOLS
Class 2 isolation equipment
CE marking in conformity with
EC directive 93/42/CEE
Manufacturer : CA-MI snc Via Ugo La Malfa n°31
43010 Pilastro (PR) Italy
Warning, consult the instruction manual
Type B equipment
Fuse
To preserve in place coolness and dry land
Conservation temperature: - 40 ÷ 70° C
~
Alternate current
Hz
Mains frequency
I
ON
0
OFF