background image

29

ENGLISH

1.2 Indication for Use

The Pulse Oximeter can be used in measuring the pulse oxygen saturation and pulse rate through 

finger. The product is suitable for being used in family, hospital, oxygen bar, community healthcare, 

physical care in sports (It can be used before or after doing sports, and it is not recommended to use the 

device during the process of having sport) and etc.

1.3 Environment Requirements

Storage Transportation Environment

a) Temperature : -40℃ ~ 60℃

b) Relative humidity : ≤95%

c) Atmospheric pressure :500 hPa~1060 hPa

Operating Environment

a) Temperature:10℃ ~ 40℃

b) Relative Humidity : ≤75%

c) Atmospheric pressure:700 hPa~1060 hPa 

1.4 Precautions

1.4.1 Attention

Point out conditions or practices that may cause damage to the device or other properties.

  Before using the device, make sure that it locates in normal working state and operating environ-

ment.

  In order to get a more accurate measurement, it should be used in a quiet and comfortable environ-

ment.

  When the device is carried from cold or hot environment to warm or humid environment, please do 

not use it immediately, wait four hours at least is recommended.

  If the device is splashed or coagulated by water, please stop operating.

  DO NOT operate the device with sharp things.

  High temperature, high pressure, gas sterilizing or immersion disinfection for the device is not per-

mitted. Refer to User Manual in the relative chapter (6.1) for cleaning and disinfection. Please turn 

off the device and disconnect it from the power.

  The device is suitable for children and adult.

  The device may not be suitable for all patients, if you can’t get a satisfactory result, please stop using 

it.

  Data averaging and signal processing generate delay on displayed and transmitted SpO

2

 value and 

alarm signal generation. The data update period is shorter than 30 s, the time for obtaining dynamic 

average values will increase, which is arisen from signal degradation, low perfusion or other interfer-

ence, it depends on the PR value. 

  The device has 3-year service life, date of manufacture: see the label.

  To further detect the alarm of individual measurement parameter, measure and check oneself or with 

a simulator, adjust the alarm limit setting and check whether the correct alarm can be triggered.

  This device has the function of alarm, patients can check on this function according to chapter 5.3 as 

a reference.

  The device has the function of limits alarm, when the measured data is beyond the highest or lowest 

limit, the device would start alarm automatically on the premise of the alarm function is on.

  The device has the function of alarm, this function can either be paused, or closed for good. This 

function could be turned on through menu operation if you need. Please check the chapter 5.3 as a 

reference.

  The maximum temperature at the SpO

2

 probe -tissue interface should be less than 41℃ which is 

measured by the temperature tester.

  During measuring, when abnormal conditions appear on the screen, please pull out your finger and 

reinsert it to measure again.

  If some unknown error appears during measuring, press “RESET” button to reset it.

  Do not contort or drag the wire of the device.

  The plethysmographic waveform is not normalized, as a signal inadequacy indicator, when it is not 

smooth and stable, the accuracy of the measured value may degrade. When it tends to be smooth 

and stable, the measured value read is the optimal and the waveform at this time is also the most 

standard. 

Summary of Contents for OXY-200

Page 1: ...MS CO LTD No 112 Qinhuang West Street Economic Technical Development Zone 066004 Qinhuangdao Hebei Province PEOPLE S REPUBLIC OF CHINA Made in China Importato da Imported by Import par Importado por G...

Page 2: ...device or replacement of the battery non detachable lithium battery can only be performed by qualified service personnel specified by manufacturer dangers such as over temperature fire or explosion m...

Page 3: ...tic field Using the device in an inappropriate environment may cause interfer ence to the surrounding radio equipment or affect its working When storing the device keep it away from children pets and...

Page 4: ...ant physiological parameter for the respiratory and circulatory system A number of diseases related to respiratory system may cause the decrease of SpO2 in the blood furthermore some other causes such...

Page 5: ...ory result please stop using it Data averaging and signal processing generate delay on displayed and transmitted SpO2 value and alarm signal generation The data update period is shorter than 30 s the...

Page 6: ...b with the blood pressure cuff arterial ductus or intra luminal tube The measured value may be inaccurate during defibrillation and in a short period after defibrillation as it has not defibrillation...

Page 7: ...a pho tosensitive element information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor 3 FUNCTIONS A SpO2 value display B Pulse rate value di...

Page 8: ...e alarm sound pause The pause time could be set by menu Long press it to confirm the alarm and there will be no alarm this time 3 Return button return to the previous menu 4 Up button down button left...

Page 9: ...the other end into computer 4 5 Structure accessories and software description A Structure main unit Oximeter probe A power supply line and A data line B Accessories one User Manual one power supply...

Page 10: ...o alarm this time 5 3 Menu operation Under the measurement interface press the menu button to enter the main menu interface as shown in Figure 5 system display clock alarm and recond etc can be set me...

Page 11: ...ss move the choice bar to the Brightness item then press left right button to set three levels B Mode move the choice bar to the Mode item then press left right button to switch display mode two kinds...

Page 12: ...e press return button to return to the main menu 5 3 4 Alarm setting When the device is turned on the red and yellow alarm lamp flash alternately once which indicates that the alarm system is working...

Page 13: ...me move the choice bar to the Alarm Volume item then press left right button to set the volume three levels d Alarm Pause sec move the choice bar to the Alarm Pause sec item then press left right to s...

Page 14: ...In Input ID interface aglimmer cursor appears in the ID input box The pink letter is the selected letter in dummy keyboard Press direction key to choose the letter to input and press menu button to i...

Page 15: ...n B Memory Manager setting In the Record menu interface move the choice bar to the Memory Manager item then press menu button to enter ID Manager interface as figure 18 Figure18 ID Manager interface I...

Page 16: ...tton or right button to move the recording time which is denoted by pink triangle 5 3 6 Close the device A In the main menu interface move the choice bar to the Power item then press menu button to cl...

Page 17: ...pls safety and monitoring performance It is recommended that the device should be inspected weekly at least When there is obvious damage stop using it B Please clean and disinfect the device before af...

Page 18: ...charged fully 2 The device works abnormally 1 Please charge the battery 2 Please contact the after sales The battery can not be full charged even after 10 hours charging time The device works abnormal...

Page 19: ...he following symbols SpO2 see note 1 Display range 0 100 Measured range 0 100 Accuracy see note 2 70 100 2 0 69 unspecified Resolution 1 PR Display range 30 bpm 250 bpm Measured range 30 bpm 250 bpm A...

Page 20: ...asurements taken over the full range By artificial inducing get the stable oxygen level to the range of 70 to 100 SpO2 compare the SpO2 values collected by the secondary standard pulse oximeter equipm...

Page 21: ...intended for use in the electromagnetic environment specified below The pur chaser or the user of the Pulse Oximeter should assure that it is used in such environment Immunity test IEC60601 test level...

Page 22: ...t 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures obj...

Page 23: ...2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 2 0 3 28 5240 5 100 5 800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 5500 5785 NOTE If necessary to achieve the...

Page 24: ...t in improper operation 4 Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the devic...