34
LCD Display
shows
abnormal
result
The cuff position was not
correct or it was not properly
tightened
Body posture was not correct
during testing
Speaking, arm or body
movement, angry, excited
or nervous during testing
Irregular heartbeat (arrhythmia)
Apply the cuff correctly and try
again
Review the “BODY POSTURE
DURING MEASUREMENT”
sections of the instructions
and re-test
Re-test when calm and without
speaking or moving during the test
It is inappropriate for people
with serious arrhythmia to use this
Electronic Sphygmomanometer
11. Troubleshooting 1
PROBLEM
POSSIBLE CAUSE
SOLUTION
ENGLISH
Summary of Contents for KD-558
Page 126: ...126 3 1 1 5 START LCD...
Page 130: ...130 15 16 15 FCC TV 17 ANSI AAMI SP 10 2002 A1 2003 A2 2006 1 a b c LCD...
Page 131: ...131 2 a LCD 2 2 1 2 2 1 2 2 b START MEM 2 2 START MEM MEM c START d 1 e 3...
Page 132: ...132 4 a velcro b 1 2 cm c d 1 2 3 4 5 5 200...
Page 133: ...133 5 a b c a b c 6 a START 6 b U1 U2 6 1 MEM 6 2 START 5 6 6 1 6 2 c 6 3 d LCD...
Page 135: ...135 7 3 7 4 LCD 7 5 7 3 7 4 7 5 d MEM 7 6 7 7 7 8 MEM 7 9 MEM 7 6 7 7 7 8 7 9 e 1 START...
Page 137: ...137 10 monitor display LCD HI Lo IEC 60601 1 8 reset display LCD 8 11 1 LCD...
Page 139: ...139 1 2 3 4 5 6 2 7 8 9 10 000 1000 10 2 75 90 BF...
Page 140: ...140 93 42 SN GIMA 12 GIMA GIMA software GIMA 0197...
Page 143: ...143 ARABIC...
Page 144: ...144 ARABIC...
Page 145: ...145 ARABIC...
Page 146: ...146 ARABIC...
Page 147: ...147 ARABIC...
Page 148: ...148 ARABIC...
Page 149: ...149 ARABIC...
Page 150: ...150 ARABIC...
Page 152: ...152 ARABIC 11...
Page 153: ...153 ARABIC 12...
Page 154: ...154 ARABIC...
Page 155: ...155 ARABIC...
Page 156: ......