GIMA SpA
DIATERMO MB202
Manuale d’istruzioni / Instruction’s Manual
MA160IGBb
20
1. INTRODUCTION
1.1 DESTINATION OF USE / SECTORS OF APPLICATION
The use of HF electro surgical equipment DIATERMO MB202 has reserved to specialized medical personnel. The equipment
has destined to a temporary use, for surgical operations in emergency room or hospital. It has foreseen its use in the monopolar
cut, cut coagulated or coagulation mode or in bipolar coagulation mode.
The equipment is proper for endoscopic operations.
The equipment is conceived for being used in the following sectors:
Causalty
Recommended
Dermatology
Recommended
Emergency
Recommended
Endoscopy
Recommended
Gastroenterology
Recommended
General surgery
Usable
Gynecology
Recommended
Neurosurgery
Recommended
Oculistcs
Recommended
Orthopedics
Usable
Otorhinolaryngology
Recommended
Pediatric surgery
Usable
Plastic surgery
Usable
Pneumology
Recommended
Urology
Recommended
Vascular surgery
Recommended
Vetenary
Recommended
1.2 STANDARD COMPOSITION
= STANDARD
code description
Pz
GMA10300.40A
DIATERMO MB 200
Unit
1
0230
Ball electrode
3
00498.04 Bipolar
adapter
1
0210 Blade
electrode
3
00100.01
Cable of feeding length 5 ms SIEMENS-IEC
1
00201.01
Handle for needles
1
MA160
Instruction Manual
1
00500.00
Kit 10 assorted electrodes
1
0220
Needle electrode
3
00500.L11
Needles for micro-surgery
1
2818C5001A
Reusable conductive rubber plate
1
2835A
Reusable handle with finger switches
1
00302.00
Watertight pedal
1
1.3 GENERAL DESCRIPTION
DIATERMO MB202
is an electro-surgical equipment suited to deliver current for monopolar cut, cut coagulated, with different
degrees of coagulation and coagulation or bipolar coagulation. The current can be delivered for the whole time of activation of
the output circuit, or for an interval of time which can be preset.
The delivery can take place once, each time the output circuit is activated, or more time spaced out with delivery time and rate
depending from the performed planning.
A total of ten different modes of use and levels of power, can be stored and recalled for the use simply. It is possible to use either
single plate neutral reference electrodes or electrodes with split conductive zone so to watch the stability of the plate to patient
impedance during the surgical intervention.
Control of the units is via front panel keys and display; mains inlet is located on the rear panel.
The units have automatic control systems that, monitoring the internal parameters, signal the possible damages/errors that are
found.
The operational parameters that are used are constantly stored so that, every time the unit is switched on or the operative method
is changed, the last utilized parameters are recalled.
The level of the emission sound can vary; every operator can choose his own level according to the noise of his working
ambient.
The units can work either through holder-handles with or without pushbuttons or through single or double foot switch command.
Moreover, applying a special adapter it is possible the unit connection to bipolar forceps.