![Gima DIATERMO MB200 Instruction Manual Download Page 31](http://html.mh-extra.com/html/gima/diatermo-mb200/diatermo-mb200_instruction-manual_2228034031.webp)
GIMA SpA
DIATERMO MB202
Manuale d’istruzioni / Instruction’s Manual
MA160IGBb
31
5. TECHNICAL CHARACTERISTICS
Tollerance Description
DIATERMO
MB202
"
Electrosurgical unit code
GMA10300.40A
0%
Minimum preselectable power
0
"
Digital level display
20%
Maximum output power CUT (W)
200
→
250
9
20%
Maximum output power CUT 90% (W)
200
→
250
9
20%
Maximum output power CUT 80% (W)
200
→
250
9
20%
Maximum output power CUT 70% (W)
200
→
250
9
20%
Maximum output power BLEND (W)
120
→
200
9
20%
Maximum output power COAG FORCED (W)
150
→
150
9
20%
Maximum output power COAG SOFT (W)
90
→
100
9
20%
Maximum output power BIPOLAR (W)
100
→
100
9
5%
Modulation factor CUT
Pure 100%
5%
Modulation factor CUT 90%
Mod. 90%
5%
Modulation factor CUT 80%
Mod. 80%
5%
Modulation factor CUT 70%
Mod. 70%
5%
Modulation factor BLEND
Pure 100%
5%
Modulation factor COAG FORCED
Mod. 60%
5%
Modulation factor COAG SOFT
Mod. 90%
5%
Modulation factor BIPOLAR (W)
Pure 100%
0.2
Crest Factor CUT
1.5
0.3
Crest Factor CUT 90%
1.6
0.3
Crest Factor CUT 80%
1.7
0.3
Crest Factor CUT 70%
1.8
0.3
Crest Factor BLEND
2.1
0.3
Crest Factor COAG FORCED
2.0
0.3
Crest Factor COAG SOFT
1.7
0.2
Crest Factor BIPOLAR (W)
1.5
10%
Working frequency
600 kHz
15%
Maximum output voltage CUT (Vpp on 5.2k
9
)
1050
15%
Maximum output voltage CUT90% (Vpp on 5.2k
9
)
1050
15%
Maximum output voltage CUT80% (Vpp on 5.2k
9
)
1050
15%
Maximum output voltage CUT70% (Vpp on 5.2k
9
)
1050
15%
Maximum output voltage BLEND (Vpp on 5.2k
9
)
1050
15%
Maximum output voltage COAG FORCED (Vpp on 5.2k
9
)
1050
15%
Maximum output voltage COAG SOFT (Vpp on 5.2k
9
)
540
15%
Maximum output voltage BIPOLAR(Vpp on 5.2k
9
)
540
0.5
Weight Kg
8
10
Size HxLxDmm
360x150x265
5%
Selectable mains power(V
~
)
115 –230
1%
Mains frequency (Hz)
50-60
0
Fuses 230V
~
(5x20) TIMED
2x 3.15A
0
Fuses 115V
~
(5x20) TIMED
2x 6.3A
10%
Electrical input power (VA)
350
10%
Electrical input current (230V
~
) (A)
1.5
10%
Electrical input current (115V
~
) (A)
3
5
Five steps adjustable sound level (from 55- to 75dBA)
"
Self-check
"
Power accuracy output warning
"
Skin Plate Electronic Control
5
5
Patient to plate contact monitoring system