GIMA SpA
DIATERMO MB202
Manuale d’istruzioni / Instruction’s Manual
MA160IGBb
27
If after having acknowledged the shown impedance the value of this it increases relatively to the acknowledged value according
to the followings increases:
for acknowledged impedance <20
9
value + 30
9
for acknowledged impedance between 20
9
and 100
9
value + 60
9
for acknowledged impedance > 100
9
value + 50%
the equipment prevents the delivery of current, it indicates the OC condition and shows the new value of impedance.
4.3.3
P
RESETTING OF THE
C
URRENT
D
ELIVERY
T
IME
The equipment can have predisposed either to deliver current all the time of activation of the output circuit or for delivering
current the time programmed in the range from 1 ms to 999 ms.
To program the delivery time, it is necessary to type in the value of time (thousandths of second) in the special panel.
The number has to be always composed of three figures and it is verifiable in the digital indicator up the keyboard. If the number
is not complete delivery it is not allowed.
If one types in the number 000 the delivery of current is interdicted.
To return to the output current delivery for the whole time of activation it is necessary to press the push button CONT.
4.3.4
R
EPETITION OF
T
IMED
D
ELIVERY
Provided that the output circuit is maintained activated, the preset time delivering can have repeated for a number of times each
second presets through the push buttons: REPEAT, UP and DOWN. The number of repetitions has shown in the display up such
push buttons. The indication stops around 3 seconds after having finished the selection: if it desires to get the indication again it
is necessary to press the REPEAT push button.
Naturally the number of repetitions per second multiplied by the number of thousandths of second for each delivery cannot
exceed the value of thousand. A wrong planning has not allowed from the equipment that automatically rectifies the number of
repetitions per second.
4.3.5
S
TORING OF
C
ONDITION OF
U
SE
Using the section STORED DATA it is possible to store 9 different operational conditions and levels of output power (program
0 is the default one).
To store the parameters it is enough to perform the following procedure:
Press the key SELECT to find the number of the program that you want to use.
Press the key STORE to store the parameters.
To recall the stored data it is enough to perform the following procedure:
Press the key SELECT to find the number of program that you want to recall. Press the key RECALL to recall the
parameters.
For modify a program it is enough to select the same program and to perform the procedure of storing once more.