4. CONNECTORS AND CONTROLS
4.1 LABEL ON THE REAR PANEL
The requirements for the safety of H.F. surgical equipment ask technical data and graphic symbols must be printed on the
cabinet or on at least one of the panels of generator unit to define its features and oversee its condition of work.
4.1.1
M
ANUFACTURER
’
S
I
DENTIFICATION
D
ATA
DIATERMO MB 200
HF electrosurgical unit is designed, manufactured and tested by the LED SpA in its own laboratories in
Aprilia (LT) - Italy.
4.1.2
T
ECHNICAL
D
ATA
Output CUT(Cut):
200 W - 250
Ω
FREQUENCY: 600
kHz
Output BLEND(Cut coagulated):
120 W - 200
Ω
FREQUENCY: 600
kHz
Output FORCED(Coagulation):
150 W - 150
Ω
FREQUENCY: 600
kHz
Output SOFT(Coagulation):
100 W - 100
Ω
FREQUENCY: 600
kHz
Output BIPOLAR (Coagulation):
100 W - 100
Ω
FREQUENCY: 600
kHz
MAIN POWER :
115/230 V - 50/60 Hz selectable
INLET
POWER:
350VA
FUSE:
(230V
~
) 2xT3.15 A (type 5x20) / (115V
~
) 2xT6.3 A (type 5x20)
DUTY - CYCLE:
10 seconds active /30 seconds pause
CLASS:
I CF
4.1.3
M
EANING OF
G
RAPHIC
S
YMBOLS
The meaning of the graphic symbols printed on DIATERMO’s cabinet is the following:
1. Floating Patient's plate: neither at low-frequency nor at high frequency earth connected
2. The equipment is protected against Cardiac Defibrillator discharge.
3. Not Ionising Radiation emitted
4. Read carefully INSTRUCTION MANUAL before to attempt the use of the equipment.
1 2 3 4