21

ENGLISH

Err09

The pulse signal is too weak or the cuff is loose.

Wrap the cuff correctly (refer to Chapter 10)

Err10

Cuff is blocked or squeezed

Wrap the cuff correctly (refer to Chapter 10)

Err14

Cuff leakage

Replace with a new cuff

Err11

Err12

Err13

The signal amplitude is too big owing to the arm or 

body moving or other reasons when measuring

Keep arm, body still, measure again

Err16

Err19

It takes too much time

CHAPTER 11 - TROUBLESHOOTING

Abnormal 

Phenomenons

Causes

Solutions

BP measurement 

values too high or 

too low.

Cuff is not connected correctly.

Correctly connect cuff. 

Talk or move arm in measurement

Keep quiet and restart a measurement.

The turnup close oppress the arm

Take off the clothes, and restart a measurement

No pressure

Cuff leakage

Buy a new cuff.

The cuff windpipe is not correctly connected 

with cuff

Correctly connect.

Cuff not inflate

Contact us.

Cuff deflate in short 

time

Loose cuff

Correctly tangle cuff.

It can not carry on measurement ,even if press the measurement button Return on the power and restart a measurement.

Abruptly turn the 

power off in adding 

pressure

No use for a long time, the batteries can be 

exhausted owing to the changed temperature

Replace all four batteries with new ones.

Hold the on/off 

button but can not 

start the device

Batteries are worn

Replace all four batteries with new ones.

The battery polarities is reversed 

Check the battery installation for proper placement of the battery 

polarities.

Cuff inflation start before press the measurement button

Stop using the device and contact us.

Cuff never deflation

Stop using the device and contact us.

Air pressure error

Deflation error

Pull out the cuff to deflate. Stop using the device and contact us.

Others 

Keep arm, body still, measure again.

No press value displayed or the value unaltered when cuff inflating

Pull out the cuff to deflate. Stop using the device and contact us.

Other phenomenon

Switch on the power once again and restart an operation.

Replace the batteries.

If no, please contact us.

CHAPTER 12 - MAINTENANCE, CLEANING AND KEEPING

*Please do obey the precautions and correct operating methods in this user manual. Otherwise, we will not responsible for any fault.

 Warning 

Remove the batteries before cleaning. The accessories and main unit must be separated for cleaning.

Maintenance is not allowed during device using.

Do not squeeze the rubber tube on the cuff.

 Caution 

High pressure disinfection to the device and accessories is not allowed.

Do not let water or cleaning agent flow into the socket to avoid device damage.

Do not soak the device and accessories in liquid.

If any damage or deterioration of the device and accessories is found, please do not use it.

Maintenance:

Clean the device and accessories regularly. It is recommended to clean them every one month. When the device or accessory gets dirty, use a dry 

and soft cloth to wipe. If they are very dirty, it is available to dip the soft cloth into water or mild detergent, and wring out, then use the cloth for 

cleaning.

The device shall  be inspected and  calibrated regularly (or according  to inspection  standard of hospital).  The inspection  can be carried out  in 

Summary of Contents for 49880

Page 1: ... ψηφιακο πιεσομετρο زاهج ةبقارم طغض مدلا يمقرلا M49880 M Rev 0 10 22 PROFESSIONAL MEDICAL PRODUCTS Contect Medical System Co Ltd Adress No 112 Qinhuang West Street Economic Technical Development Zone Qinhuangdao Hebei Province People s Republic of China Made in China CONTEC08E GIMA 49880 Shanghai International Holding Corp Gmbh Europe Address Eiffestrasse 80 20537 Hamburg Germany Gima S p A Via Ma...

Page 2: ...oduct to our company Responsibility for our company Our company have the responsibility to provide qualified product which conform to company standard of this product Our company will provide the circuit diagram calibration method and other information at the request of the user to help the appropriate and qualified technicians to repair those parts designated by our company Our company have the r...

Page 3: ...nd the person who can t express oneself please use the device under the guidance of a doctor Otherwise it may cause accident or dissension Please do not use for any other purpose except BP measurement Otherwise it may cause accident or holdback Please use special cuff Otherwise it is possible that measurement result is incorrect Please do not keep the cuff in the over inflated state for a long tim...

Page 4: ...e manganese or alkaline batteries do not use batteries of other types Otherwise it may cause fire Do not mix old and new batteries and batteries of different types Otherwise it may cause battery leakage heat rupture and damage to Electronic Sphygmomanometer Please don t put wrong the positive and negative of battery When the batteries power exhausts replace with four new batteries at the same time...

Page 5: ...nd digital equipment connected to this device must be certified to IEC standards such as IEC60950 Information technology equipment Safety and IEC60601 1 Medical electrical equipment Safety and all equipment should be connected to in accordance with the requirement of the valid version of the IEC60601 1 1 system standard The person connecting the additional equipment to the signal input and output ...

Page 6: ...ur legs Advice Try to measure your blood pressure at the same time each day with the same arm and the same pose for consistency The high and low location of cuff will cause changes in measure results Do not touch the sphygmomanometer cuff and windpipe during measure Measurements should be taken in a quiet place and the body relax Remain still 4 5 minutes before measurement Do not talk and movement...

Page 7: ...ay be dangerous Follow the instructions of your physician Note Wait at least 4 5 minutes between measurements When repeatedly measuring the accurate blood pressure value may not be measured due to congestion in the arm Please measure after the blood flow is smooth When the screen displays Err the measure can t be carried out correctly Irregular pulse icon is displayed in the measurement results if...

Page 8: ...er Medical Device complies with Directive 93 42 EEC Authorized representative in the European community Class II applied Socket for power adapter Interface for connecting cuff Voice deactivation Voice activation Great movement when measuring Cuff tied correctly Artery indicator label CHAPTER 10 ERROR MESSAGE When the high pressure position appears Err and the low pressure position appears the erro...

Page 9: ...ion start before press the measurement button Stop using the device and contact us Cuff never deflation Stop using the device and contact us Air pressure error Deflation error Pull out the cuff to deflate Stop using the device and contact us Others Keep arm body still measure again No press value displayed or the value unaltered when cuff inflating Pull out the cuff to deflate Stop using the devic...

Page 10: ...hod Oscillometric method Working mode Automatic Operation mode Continuous operation Pressure Range 0 297 mmHg 0 39 6 kPa Measurement range Pressure SYS 30 270 mmHg 4 36 kPa DIA 10 220 mmHg 1 3 29 3 kPa Pulse 40 240 min Inflation 160 5 mmHg 21 33 0 67 kPa Overpressure protect 297 3 mmHg 39 6 0 4 kPa Resolution Pressure 1 mmHg 0 133 kPa Accuracy Static pressure 3 mmHg 0 4 kPa Error The BP value meas...

Page 11: ...elow The customer or the user of the device should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at l...

Page 12: ...ixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is ...

Page 13: ... the pins of connectors marked with an ESD warning symbol unless electrostatic discharge precautions are used the device should not connect to these connectors In order to avoid the accumulation of electrostatic charge it is recommended to store maintain and use the equipment at a relative humidity of 30 or more The floor should be covered with ESD dissipated carpets or similar materials In the us...

Reviews: