background image

24

ENGLISH

Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should 

assure that it is used in such an environment.

Immunity 

test

IEC 60601 test level

Compliance 

level

Electromagnetic environment - guidance

Portable  and  mobile  RF  communications  equipment  should  be 

used no closer to any part of the 

device

, including cables, than the 

recommended  separation  distance  calculated  from  the  equation 

applicable to the frequency of the transmitter.

Conducted RF

IEC61000-4-6

Radiated RF

IEC61000-4-3

3 V/m

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 V/m

3 V/m

Recommended separation distance

P

E

d

=

1

5

.

3

P

V

d

=

1

5

.

3

P

E

d

=

1

7

80 MHz to 800 MHz

800 MHz to 2.5 GHz

Where 

P

 is the maximum output power rating of the transmitter in 

watts (W) according to the transmitter manufacturer and 

d

 is the 

recommended separation distance in metres (m).

Field  strengths  from  fixed  RF  transmitters,  as  determined  by  an 

electromagnetic site survey,

a

 should be less than the compliance 

level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with 

the following symbol:

NOTE 1 

At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 

These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people.

A Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 

amateur  radio,  AM  and  FM  radio  broadcast  and  TV  broadcast  cannot  be  predicted  theoretically  with  accuracy.  To  assess  the 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 

field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should 

be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as 

reorienting or relocating the device.

B Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for 

EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF communications equipment and the 

device

The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. 

The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum 

distance between portable and mobile RF communications equipment (transmitters) and the device as recommended 

below, according to the maximum output power of the communications equipment.

Rated 

maximum 

output power of 

transmitter

(W)

Separation distance according to frequency of transmitter

(m)

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.17

1.17

2.33

10

3.69

3.69

7.38

100

11.67

11.67

23.33

Summary of Contents for 49880

Page 1: ... ψηφιακο πιεσομετρο زاهج ةبقارم طغض مدلا يمقرلا M49880 M Rev 0 10 22 PROFESSIONAL MEDICAL PRODUCTS Contect Medical System Co Ltd Adress No 112 Qinhuang West Street Economic Technical Development Zone Qinhuangdao Hebei Province People s Republic of China Made in China CONTEC08E GIMA 49880 Shanghai International Holding Corp Gmbh Europe Address Eiffestrasse 80 20537 Hamburg Germany Gima S p A Via Ma...

Page 2: ...oduct to our company Responsibility for our company Our company have the responsibility to provide qualified product which conform to company standard of this product Our company will provide the circuit diagram calibration method and other information at the request of the user to help the appropriate and qualified technicians to repair those parts designated by our company Our company have the r...

Page 3: ...nd the person who can t express oneself please use the device under the guidance of a doctor Otherwise it may cause accident or dissension Please do not use for any other purpose except BP measurement Otherwise it may cause accident or holdback Please use special cuff Otherwise it is possible that measurement result is incorrect Please do not keep the cuff in the over inflated state for a long tim...

Page 4: ...e manganese or alkaline batteries do not use batteries of other types Otherwise it may cause fire Do not mix old and new batteries and batteries of different types Otherwise it may cause battery leakage heat rupture and damage to Electronic Sphygmomanometer Please don t put wrong the positive and negative of battery When the batteries power exhausts replace with four new batteries at the same time...

Page 5: ...nd digital equipment connected to this device must be certified to IEC standards such as IEC60950 Information technology equipment Safety and IEC60601 1 Medical electrical equipment Safety and all equipment should be connected to in accordance with the requirement of the valid version of the IEC60601 1 1 system standard The person connecting the additional equipment to the signal input and output ...

Page 6: ...ur legs Advice Try to measure your blood pressure at the same time each day with the same arm and the same pose for consistency The high and low location of cuff will cause changes in measure results Do not touch the sphygmomanometer cuff and windpipe during measure Measurements should be taken in a quiet place and the body relax Remain still 4 5 minutes before measurement Do not talk and movement...

Page 7: ...ay be dangerous Follow the instructions of your physician Note Wait at least 4 5 minutes between measurements When repeatedly measuring the accurate blood pressure value may not be measured due to congestion in the arm Please measure after the blood flow is smooth When the screen displays Err the measure can t be carried out correctly Irregular pulse icon is displayed in the measurement results if...

Page 8: ...er Medical Device complies with Directive 93 42 EEC Authorized representative in the European community Class II applied Socket for power adapter Interface for connecting cuff Voice deactivation Voice activation Great movement when measuring Cuff tied correctly Artery indicator label CHAPTER 10 ERROR MESSAGE When the high pressure position appears Err and the low pressure position appears the erro...

Page 9: ...ion start before press the measurement button Stop using the device and contact us Cuff never deflation Stop using the device and contact us Air pressure error Deflation error Pull out the cuff to deflate Stop using the device and contact us Others Keep arm body still measure again No press value displayed or the value unaltered when cuff inflating Pull out the cuff to deflate Stop using the devic...

Page 10: ...hod Oscillometric method Working mode Automatic Operation mode Continuous operation Pressure Range 0 297 mmHg 0 39 6 kPa Measurement range Pressure SYS 30 270 mmHg 4 36 kPa DIA 10 220 mmHg 1 3 29 3 kPa Pulse 40 240 min Inflation 160 5 mmHg 21 33 0 67 kPa Overpressure protect 297 3 mmHg 39 6 0 4 kPa Resolution Pressure 1 mmHg 0 133 kPa Accuracy Static pressure 3 mmHg 0 4 kPa Error The BP value meas...

Page 11: ...elow The customer or the user of the device should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at l...

Page 12: ...ixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is ...

Page 13: ... the pins of connectors marked with an ESD warning symbol unless electrostatic discharge precautions are used the device should not connect to these connectors In order to avoid the accumulation of electrostatic charge it is recommended to store maintain and use the equipment at a relative humidity of 30 or more The floor should be covered with ESD dissipated carpets or similar materials In the us...

Reviews: