When cut off the power supply, first cut off the connection of power socket and the regulated power supply, then cut off the connection of regulated 

power supply and the 

s

phygmomanometer.

Please be sure to use dedicated medical grade power adapter.

 Note 

When regulated power supply and batteries are both used at the same time, the battery power will not be consumed.

Switch regulated power supply and battery as power supply when the device is off, otherwise, the device may shutdown due to power failure.

The device can be used normally after it is turned on ,without waiting for the device to be prepared.

CHAPTER 6 - BUTTON FUNCTIONS

6.1 Description for button operation 

All the operations to the Electronic Sphygmomanometer are through buttons. The names of the buttons are above them. They are:

 

Left button is “M” button, under “OFF” state, press this button to enter the review interface (refer to Chapter 8 for details.).

 

Right button is “START/STOP” button, under “OFF” state, press this button to enter measurement mode, inflate the cuff to measure blood 

pressure, press this button again to turn off the device.

6.2 Units setting 

Under “OFF” state, press “M” button and “START/STOP” button simultaneously for 5 s to enter the setting interface, the default unit in this interface is 

“mmHg”; short press “M” button to switch the unit between “mmHg” and “kPa”.

6.3 Volume setting (optional for devices with voice function)

 

Press “START/STOP” button again in the unit setting interface to enter the volume setting interface. Press “M” button to change the volume, the 

maximum volume is 4, and the minimum is 0 (silence).

 

After completing the setting, repeatedly press the “START/STOP” button to turn the device off.

 Note 

 

The default unit of the device when leaving factory is mmHg.

 

In the volume setting interface, press “START/STOP” button to enter the factory setting interface, in which the “CAL” is the static pressure interface, 

and the “FAC” is the aging interface, which does not require user to operate. If you want to end the interface, press the “START/STOP” button 

twice to turn the device off.

CHAPTER 7 - THE USAGE METHOD OF SPHYGMOMANOMETER

7.1 Accurate Measurement Way

Measurement in quiet and relaxing state.

1.  Adopt a comfortable sitting position, use back and arms to support the body.

2.  Place your elbow on a table, the palm faces up and the body is relaxed.

3.  The cuff is level with your heart.

4.  Feet flat on the floor, and do not cross your legs.

 

Advice 

 

Try to measure your blood pressure at the same time each day with the same arm and the same pose for consistency.

The high and low location of cuff will cause changes in measure results.

Do not touch the sphygmomanometer, cuff and windpipe during measure.  

Measurements should be taken in a quiet place and the body relax.

Remain still 4~5 minutes before measurement.

Do not talk and movement during the measurement. Relax the body, do not let the muscle activity.

Wait 4~5 minutes between measurements.

Do not use precision instrument near the Sphygmomanometer.

 

Warning 

When repeatedly measuring, the accurate blood pressure value may not be measured due to congestion in the arm

Please measure after the blood 

flow is smooth.

Repeated measurement for a long period of time, limbs rubbing with the cuff may be accompanied by purpura, ischemia and nerve damage. When 

m

easurement a patient, it is necessary to frequently check the color, warmth and sensitivity of the distal of the limb. Once any abnormalities are 

observed, place the cuff in another position or stop the blood pressure measurement immediately.

Please use the device at an environment of suitable temperature and humidity  otherwise it will cause measurement error.

Do not twist or wrap the airway tube. It can cause constant pressure in the cuff which can block blood flow and cause serious damage to the patient.

 

Do not use the cuff on the injured area, which will cause more serious damage to the area.

 

Do not use the cuff in the area where the treatment is being performed inside blood vessel or the arteriovenous connection. This may cause 

temporary blockage of blood flow and cause injury to the patient.

 

Do not use the cuff on the side of the mastectomy.

 

When using the cuff to pressurize, some of the body’s functions may temporarily weaken. Do not use the measurement medical electrical equipment 

at the appropriate arm position.

 

Do not move during measurement, it will have a delayed effect on the patient’s blood flow.

The device need to be placed for 2 hours from the minimum storage temperature to being ready for its intended use.

The device need to be placed for 4 hours from the highest storage temperature to being ready for its intended use.

 Note 

The following conditions may also cause changes in the blood pressure measurement value.

Take the measurement in one hour after meal or after drinking alcohol, coffee or after smoking, exercise, bathing;

Using incorrect posture such as standing or lying down, etc;

The patient speak or move his body during measurement;

When measuring, the patient is nervous, excited, emotional instability;

Summary of Contents for 49880

Page 1: ... ψηφιακο πιεσομετρο زاهج ةبقارم طغض مدلا يمقرلا M49880 M Rev 0 10 22 PROFESSIONAL MEDICAL PRODUCTS Contect Medical System Co Ltd Adress No 112 Qinhuang West Street Economic Technical Development Zone Qinhuangdao Hebei Province People s Republic of China Made in China CONTEC08E GIMA 49880 Shanghai International Holding Corp Gmbh Europe Address Eiffestrasse 80 20537 Hamburg Germany Gima S p A Via Ma...

Page 2: ...oduct to our company Responsibility for our company Our company have the responsibility to provide qualified product which conform to company standard of this product Our company will provide the circuit diagram calibration method and other information at the request of the user to help the appropriate and qualified technicians to repair those parts designated by our company Our company have the r...

Page 3: ...nd the person who can t express oneself please use the device under the guidance of a doctor Otherwise it may cause accident or dissension Please do not use for any other purpose except BP measurement Otherwise it may cause accident or holdback Please use special cuff Otherwise it is possible that measurement result is incorrect Please do not keep the cuff in the over inflated state for a long tim...

Page 4: ...e manganese or alkaline batteries do not use batteries of other types Otherwise it may cause fire Do not mix old and new batteries and batteries of different types Otherwise it may cause battery leakage heat rupture and damage to Electronic Sphygmomanometer Please don t put wrong the positive and negative of battery When the batteries power exhausts replace with four new batteries at the same time...

Page 5: ...nd digital equipment connected to this device must be certified to IEC standards such as IEC60950 Information technology equipment Safety and IEC60601 1 Medical electrical equipment Safety and all equipment should be connected to in accordance with the requirement of the valid version of the IEC60601 1 1 system standard The person connecting the additional equipment to the signal input and output ...

Page 6: ...ur legs Advice Try to measure your blood pressure at the same time each day with the same arm and the same pose for consistency The high and low location of cuff will cause changes in measure results Do not touch the sphygmomanometer cuff and windpipe during measure Measurements should be taken in a quiet place and the body relax Remain still 4 5 minutes before measurement Do not talk and movement...

Page 7: ...ay be dangerous Follow the instructions of your physician Note Wait at least 4 5 minutes between measurements When repeatedly measuring the accurate blood pressure value may not be measured due to congestion in the arm Please measure after the blood flow is smooth When the screen displays Err the measure can t be carried out correctly Irregular pulse icon is displayed in the measurement results if...

Page 8: ...er Medical Device complies with Directive 93 42 EEC Authorized representative in the European community Class II applied Socket for power adapter Interface for connecting cuff Voice deactivation Voice activation Great movement when measuring Cuff tied correctly Artery indicator label CHAPTER 10 ERROR MESSAGE When the high pressure position appears Err and the low pressure position appears the erro...

Page 9: ...ion start before press the measurement button Stop using the device and contact us Cuff never deflation Stop using the device and contact us Air pressure error Deflation error Pull out the cuff to deflate Stop using the device and contact us Others Keep arm body still measure again No press value displayed or the value unaltered when cuff inflating Pull out the cuff to deflate Stop using the devic...

Page 10: ...hod Oscillometric method Working mode Automatic Operation mode Continuous operation Pressure Range 0 297 mmHg 0 39 6 kPa Measurement range Pressure SYS 30 270 mmHg 4 36 kPa DIA 10 220 mmHg 1 3 29 3 kPa Pulse 40 240 min Inflation 160 5 mmHg 21 33 0 67 kPa Overpressure protect 297 3 mmHg 39 6 0 4 kPa Resolution Pressure 1 mmHg 0 133 kPa Accuracy Static pressure 3 mmHg 0 4 kPa Error The BP value meas...

Page 11: ...elow The customer or the user of the device should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at l...

Page 12: ...ixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is ...

Page 13: ... the pins of connectors marked with an ESD warning symbol unless electrostatic discharge precautions are used the device should not connect to these connectors In order to avoid the accumulation of electrostatic charge it is recommended to store maintain and use the equipment at a relative humidity of 30 or more The floor should be covered with ESD dissipated carpets or similar materials In the us...

Reviews: