19

ENGLISH

The room temperature rise or fall sharply, or the environment of measurement often changes;

Measuring in a moving vehicle;

The high and low location of cuff will cause changes in measurement results;

Continuous measurement for a long time.

7.2 Applying the Cuff

Both left and right arm can be measured.

Bare your arm or cloth close-fitting clothing during measurement.

Carry out the operation in a room with comfortable temperature.

When measuring, take the thick clothes off instead of rolling up the sleeves.

In order to measure accurately, pay attention to applying the cuff properly (left arm).

① Insert the arm cuff air plug in the cuff socket of sphygmomanometer

.

② Stretch cuff into a barrel for the arm can conformable enter into the barrel

③ Left arm penetrate through the cuff, the air tube of the cuff will pass the top of your palm.

④ Wrap the cuff to your upper arm. Make the air tube inside the forearm and aligned with your middle finger.

⑤ The bottom of the cuff should be approximately 2cm~3cm above your elbow.

⑥ Be fixed with cloths, and wrapped tight cuff, the arm and the cuff should not have gaps.

7.3 Measurement BP

①Under “OFF” state, press “START/STOP” button to start measuring.

During measurement, please keep correct pose and quiet state, the body could not move.The “Movement” icon appears if patient moves, and continue 

measuring may lead to inaccurate measurement.

If you want to abort the measurement 

Press

START/STOP

button, 

t

he device will stop inflating, 

and

 release the air from the cuff

.

②Display the measurement results after finishing measuring.The pressure bar on the right side visually demonstrates the pressure level.

③Confirm the Measurement Value

The World Health Organization has established globally accepted standards for the assessment of high blood pressure readings.The SYS higher than 

135mmHg or the DIA lower than 85mmHg are used as the criteria of hypertension, and the pressure bar at the right side lights up in red.

*Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician.

 Note 

Wait at least 4-5 minutes between measurements.

When repeatedly measuring, the accurate blood pressure value may not be measured due to congestion in the arm

Please measure after the 

blood flow is smooth.

When the screen displays Err, the measure can’t be carried out correctly.

Irregular pulse icon is displayed in the measurement results if the pulse internal is irregular during measuring, which may cause it is unable to take 

measurement correctly. Please keep quiet and remeasure. If the irregular pulse icon appears frequently, please consult a doctor.

The minimum value of the patient’s physiological signal is the minimum limit that the device can measure. The device may obtain inaccurate 

measurement results when it is operated below the minimum amplitude or minimum value of the patient’s physiological signal.

*The device will automatically turn off after five minutes in which there is no operation to the device, even if you forget to turn the power off.

CHAPTER 8 - MEMORY FUNCTION

The device can store NIBP values automatically, display up to 199 set of measurement results.

If 199 set of measurement data have been stored in current device, when saving the 200th set of data, the earliest set of data will be overwritten. If no 

measurement values, the memory values can be not numerated.

Memory function can not be used during measuring.

When there is no measurement values, “

---

” will display on the review interface.

8.1 Review the Memory Value

1.Under “OFF” state, press “M” button to display the average value of the latest three set of data, when the number of measurement data is less than 

three groups, it will supplement automatically. Continue to press “M” button in current interface to view all measurement records.

8.2 Delete Memory Values

1.Users can delete all memory values of the current user instead of separately delete one memory value

2.Under the memory interface, press “M” button and “START/STOP” button simultaneously for more than 5 s, after “DEL” appears on the screen, all 

memory values will be deleted.

 Caution 

When querying the measurement records, please press “M” button continuously to query one by one.

CHAPTER 9 - KEY 

A

ND SYMBOLS

Summary of Contents for 49880

Page 1: ... ψηφιακο πιεσομετρο زاهج ةبقارم طغض مدلا يمقرلا M49880 M Rev 0 10 22 PROFESSIONAL MEDICAL PRODUCTS Contect Medical System Co Ltd Adress No 112 Qinhuang West Street Economic Technical Development Zone Qinhuangdao Hebei Province People s Republic of China Made in China CONTEC08E GIMA 49880 Shanghai International Holding Corp Gmbh Europe Address Eiffestrasse 80 20537 Hamburg Germany Gima S p A Via Ma...

Page 2: ...oduct to our company Responsibility for our company Our company have the responsibility to provide qualified product which conform to company standard of this product Our company will provide the circuit diagram calibration method and other information at the request of the user to help the appropriate and qualified technicians to repair those parts designated by our company Our company have the r...

Page 3: ...nd the person who can t express oneself please use the device under the guidance of a doctor Otherwise it may cause accident or dissension Please do not use for any other purpose except BP measurement Otherwise it may cause accident or holdback Please use special cuff Otherwise it is possible that measurement result is incorrect Please do not keep the cuff in the over inflated state for a long tim...

Page 4: ...e manganese or alkaline batteries do not use batteries of other types Otherwise it may cause fire Do not mix old and new batteries and batteries of different types Otherwise it may cause battery leakage heat rupture and damage to Electronic Sphygmomanometer Please don t put wrong the positive and negative of battery When the batteries power exhausts replace with four new batteries at the same time...

Page 5: ...nd digital equipment connected to this device must be certified to IEC standards such as IEC60950 Information technology equipment Safety and IEC60601 1 Medical electrical equipment Safety and all equipment should be connected to in accordance with the requirement of the valid version of the IEC60601 1 1 system standard The person connecting the additional equipment to the signal input and output ...

Page 6: ...ur legs Advice Try to measure your blood pressure at the same time each day with the same arm and the same pose for consistency The high and low location of cuff will cause changes in measure results Do not touch the sphygmomanometer cuff and windpipe during measure Measurements should be taken in a quiet place and the body relax Remain still 4 5 minutes before measurement Do not talk and movement...

Page 7: ...ay be dangerous Follow the instructions of your physician Note Wait at least 4 5 minutes between measurements When repeatedly measuring the accurate blood pressure value may not be measured due to congestion in the arm Please measure after the blood flow is smooth When the screen displays Err the measure can t be carried out correctly Irregular pulse icon is displayed in the measurement results if...

Page 8: ...er Medical Device complies with Directive 93 42 EEC Authorized representative in the European community Class II applied Socket for power adapter Interface for connecting cuff Voice deactivation Voice activation Great movement when measuring Cuff tied correctly Artery indicator label CHAPTER 10 ERROR MESSAGE When the high pressure position appears Err and the low pressure position appears the erro...

Page 9: ...ion start before press the measurement button Stop using the device and contact us Cuff never deflation Stop using the device and contact us Air pressure error Deflation error Pull out the cuff to deflate Stop using the device and contact us Others Keep arm body still measure again No press value displayed or the value unaltered when cuff inflating Pull out the cuff to deflate Stop using the devic...

Page 10: ...hod Oscillometric method Working mode Automatic Operation mode Continuous operation Pressure Range 0 297 mmHg 0 39 6 kPa Measurement range Pressure SYS 30 270 mmHg 4 36 kPa DIA 10 220 mmHg 1 3 29 3 kPa Pulse 40 240 min Inflation 160 5 mmHg 21 33 0 67 kPa Overpressure protect 297 3 mmHg 39 6 0 4 kPa Resolution Pressure 1 mmHg 0 133 kPa Accuracy Static pressure 3 mmHg 0 4 kPa Error The BP value meas...

Page 11: ...elow The customer or the user of the device should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floor are covered with synthetic material the relative humidity should be at l...

Page 12: ...ixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the device is ...

Page 13: ... the pins of connectors marked with an ESD warning symbol unless electrostatic discharge precautions are used the device should not connect to these connectors In order to avoid the accumulation of electrostatic charge it is recommended to store maintain and use the equipment at a relative humidity of 30 or more The floor should be covered with ESD dissipated carpets or similar materials In the us...

Reviews: