Guidance and manufacture’s declaration-electromagnetic immunity
The device is intended for use in the electromagnetic environment listed below, and should only be 

used in such environments:
Immunity test

EN 60601 

test level

Compliance 

level

Electromagnetic 

environment-guidance

Electrostatic 

discharge (ESD) 

IEC 61000-4-2

±6kV contact 

±8kV air

±6kV contact 

±8kV air

Floors  should  be  wood,  concrete  or 

ceramic  tile.  If  floor  are  covered  with 

synthetic material, the relative humidity 

should be at least 30%.

Power frequency

magnetic field 

IEC 61000-4-8

3A/m

3A/m

Power frequency magnetic fields should 

be at levels characteristic of a typical lo-

cation in a typical commercial or hospi-

tal environment.

Electrical fast 

transient/burst

IEC 61000-4-4

±2kV for power 

supply lines

±2kV for power 

supply lines 

Mains power quality should be that of a 

typical commercial or hospital environ-

ment.

Surge 

IEC 61000-4-5

±1kV line(s)

to line(s)

±1kV line(s)

to line(s)

Mains power quality should be that of a 

typical commercial or hospital environ-

ment.

Interruptions and 

voltage variations

on power supply

input lines

IEC 61000-4-11

<5% U

T

 

(>95% dip in U

T

for 0.5 cycle 

40% U

T

 

(60% dip in U

T

for 5 cycles 

70% U

T

 

(30% dip in U

T

for 25 cycles

<5% U

T

 

(>95% dip in U

T

for 5sec 

<5% U

T

 

(>95% dip in U

T

for 0.5 cycle 

40% U

T

 

(60% dip in U

T

for 5 cycles 

70% U

T

 

(30% dip in U

T

for 25 cycles

<5% U

T

 

(>95% dip in U

T

for 5sec

Mains power quality should be that of a 

typical commercial or hospital environ-

ment.

Summary of Contents for 32803

Page 1: ...nte questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et...

Page 2: ...Bouton M1 10 Bouton M2 11 Bouton START 12 tui 13 4 piles AA 14 Adaptateur AC 1 Hauptstruktur 2 Display 3 Luftanschlussbuchse 4 Schlauchstecker 5 Luftschlauch 6 Armmanschette 7 D Ring 8 Taste F 9 Taste...

Page 3: ...nstruction for patients with cardio vascular diseases Under no circumstances should you alter the dosages of any drugs prescribed by your doctor 3 Accurate measurement of blood pressure may be difficu...

Page 4: ...least 600 mA Output plug polarity inner External diameter 5 5 0 1 mm Internal diameter 2 1 0 1 mm Length 10 0 3 mm CORRECT POSTURE FOR MEASUREMENT 1 Sit at the table and let the table support your arm...

Page 5: ...o get accurate date and time the user should preset the date and time correctly before the first use of this device The operation procedure for presetting Date Time is as follows 1 When the device is...

Page 6: ...hich means the alarm 01 is removed 2 When the device stands by press button F three and four times respectively to enter into alarm 02 and 03 mode Repeat the above process to remove the alarm 02 and a...

Page 7: ...able pressure level to ensure a successful measurement It is not a fault Rapid deflation during measurement If you do not feel well during measurement or want to stop the measurement for some reason y...

Page 8: ...sary to protect this device against high moisture direct sunlight shock solvent alcohol and gasoline 2 Remove the batteries if the device is to be stored for a long time and keep the batteries far fro...

Page 9: ...RT button is pres sed The batteries have run down The polarity of battery is wrong The contact of battery compart ment is polluted Replace all the batteries with new ones Install the batteries correct...

Page 10: ...both as regards the material and the production The warranty is valid for 12 months from the date of supply of GIMA During the period of validity of the warranty GIMA will repair and or replace free...

Page 11: ...ic of a typical lo cation in a typical commercial or hospi tal environment Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 2kV for power supply lines Mains power quality shoul...

Page 12: ...a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800 MHz the higher freque...

Page 13: ......

Page 14: ......

Page 15: ......

Reviews: