Guidance and manufacture’s declaration-electromagnetic immunity
The device is intended for use in the electromagnetic environment listed below, and should only be 

used in such environments:
Immunity test

IEC 60601 

test level

Compliance 

level

Electromagnetic environment - 

guidance

Conducted RF 

IEC61000-4-6

Radiated RF 

IEC61000-4-3

3Vrms

150KHz to 80MHz 

10V/m

80MHz to 2.5GHz

3Vrms

3V/m

Portable and mobile RF communica-

tions equipment should be used no clo-

ser to any part of the device, including 

cables, than the recommended separa-

tion distance calculated from the equa-

tion applicable to the frequency 

of the transmitter.

Recommended separation distance

 

                        

80MHz to 800MHz

                        

800MHz to 2.5GHz

Where 

P

 is the maximum output power 

rating of the transmitter in watts (W) ac-

cording to the transmitter manufacturer 

and 

d

 is the recommended separation 

distance in meters (m).

Field strengths from fixed RF transmit-

ters,  as  determined  by  an  electroma-

gnetic site survey,

a

 should be less than 

the compliance level in each frequency 

range.

b

Interference may occur in the vicinity of 

equipment marked with 

the following symbol:

NOTE 1 At 80MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 

absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telepho-

nes and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot 

be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 

transmitters, an electromagnetic site survey should be considered. If the measured field strength in 

the location in which the device is used exceeds themapplicable RF compliance level above, the de-

vice should be observed to verify normal operation. If abnormal performance is observed, additional 

measures may be necessary, such as reorienting or relocating the device.

Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.

3.5

V

 

1

√P

d

=

3.5

E

 

1

√P

d=

 7

E

 

1

√P

d=

Summary of Contents for 32803

Page 1: ...nte questo manuale prima di utilizzare il prodotto ATTENTION The operators must carefully read and completely understand the present manual before using the product AVIS Les op rateurs doivent lire et...

Page 2: ...Bouton M1 10 Bouton M2 11 Bouton START 12 tui 13 4 piles AA 14 Adaptateur AC 1 Hauptstruktur 2 Display 3 Luftanschlussbuchse 4 Schlauchstecker 5 Luftschlauch 6 Armmanschette 7 D Ring 8 Taste F 9 Taste...

Page 3: ...nstruction for patients with cardio vascular diseases Under no circumstances should you alter the dosages of any drugs prescribed by your doctor 3 Accurate measurement of blood pressure may be difficu...

Page 4: ...least 600 mA Output plug polarity inner External diameter 5 5 0 1 mm Internal diameter 2 1 0 1 mm Length 10 0 3 mm CORRECT POSTURE FOR MEASUREMENT 1 Sit at the table and let the table support your arm...

Page 5: ...o get accurate date and time the user should preset the date and time correctly before the first use of this device The operation procedure for presetting Date Time is as follows 1 When the device is...

Page 6: ...hich means the alarm 01 is removed 2 When the device stands by press button F three and four times respectively to enter into alarm 02 and 03 mode Repeat the above process to remove the alarm 02 and a...

Page 7: ...able pressure level to ensure a successful measurement It is not a fault Rapid deflation during measurement If you do not feel well during measurement or want to stop the measurement for some reason y...

Page 8: ...sary to protect this device against high moisture direct sunlight shock solvent alcohol and gasoline 2 Remove the batteries if the device is to be stored for a long time and keep the batteries far fro...

Page 9: ...RT button is pres sed The batteries have run down The polarity of battery is wrong The contact of battery compart ment is polluted Replace all the batteries with new ones Install the batteries correct...

Page 10: ...both as regards the material and the production The warranty is valid for 12 months from the date of supply of GIMA During the period of validity of the warranty GIMA will repair and or replace free...

Page 11: ...ic of a typical lo cation in a typical commercial or hospi tal environment Electrical fast transient burst IEC 61000 4 4 2kV for power supply lines 2kV for power supply lines Mains power quality shoul...

Page 12: ...a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800 MHz the higher freque...

Page 13: ......

Page 14: ......

Page 15: ......

Reviews: