Pipetman P2 and P10 are delivered with the
adaptor in place and configured to accept
DL10 tips. When you use D10 tips, which are
shorter, you must reposition the adaptor as
follows.
1) Pull the adaptor away from the metallic rod.
2) Turn the adaptor through 180°.
3) Refit the adaptor so that the end of the
metallic rod engages the shorter slot of the
adaptor.
4) Finally, check that end of the metallic rod is
correctly seated in the appropriate slot of
the adaptor; the ‘dimple’ on the metallic rod
must be engaged in the corresponding
hole on the adaptor.
9 - GLP FEATURES
These are as follows:
• Serial Number:
engraved on body
of the pipette.
• Bar Code:
on the box and with
the certificate (can
be transferred).
• ID Tag (Application
or User).
10 - TROUBLESHOOTING
You may be able to identify and to correct the
problem by reference to the following table.
If you can’t solve the problem, contact your
Gilson representative.
However, a quick inspection of the pipette may
help you isolate the problem.
ID Tag