19
ENGLISH
The piston assembly must not be
autoclaved.
1) Remove the tip-ejector (see above).
2) Unscrew the connecting nut by turning it
counterclockwise, by hand.
3) Carefully separate the lower and upper
parts.
4) Remove the piston assembly, O-ring and
seal.
5) Clean and decontaminate the piston
assembly.
6
Reassemble the pipette (refer to the figure).
7) Tighten the connecting nut (turn by hand,
clockwise).
8) Refit the tip-ejector (see above).
Changing the O-ring
The O-ring and seal are normally to be found
on the piston; they must not be autoclaved, if
worn or damaged in any way (chemical or
mechanical), they must be replaced. The di-
mensions of the O-ring vary according to the
model of pipette.
1) Remove the tip-ejector (see above).
2) Unscrew the connecting nut by turning it
counterclockwise, by hand.
3) Carefully separate the lower and upper parts.
4) Remove the piston assembly, O-ring and
seal.
5) Clean or replace the seal then the O-ring.
6) Reassemble the pipette (refer to the figure
opposite).
7) Tighten the connecting nut (turn by hand,
clockwise).
8) Refit the tip-ejector (see above).