INTRODUCTION
|
PIPETMAN® CLASSIC
Maximum Permissible Errors
The data provided in the following tables
conform to the ISO 8655-2 standard. With
a precise pipetting technique (refer to
on page 9), the
P2 model may be used to aspirate volumes as low
as 0.1 µL and the P10 model as low as 0.5 µL.
Figure 2
PIPETMAN® Classic single channel model
VOLUME
(µL)
GILSON
ISO 8655
SYSTEMATIC
ERROR (µL)
RANDOM
ERROR (µL)
SYSTEMATIC
ERROR (µL)
RANDOM
ERROR (µL)
P2 (P/N F144801)
WITH D10 OR DL10 TIPS
0.2
± 0.024
≤
0.012
± 0.08
≤
0.04
0.5
± 0.025
≤
0.012
± 0.08
≤
0.04
1
± 0.025
≤
0.012
± 0.08
≤
0.04
2
± 0.030
≤
0.014
± 0.08
≤
0.04
P10 (P/N F144802)
WITH D10 OR DL10 TIPS
1
± 0.025
≤
0.012
± 0.12
≤
0.08
5
± 0.075
≤
0.030
± 0.12
≤
0.08
10
± 0.100
≤
0.040
± 0.12
≤
0.08
P20 (P/N F123600)
WITH D200 TIPS
2
± 0.10
≤
0.030
± 0.2
≤
0.1
5
± 0.10
≤
0.040
± 0.2
≤
0.1
10
± 0.10
≤
0.050
± 0.2
≤
0.1
20
± 0.20
≤
0.060
± 0.2
≤
0.1
P100 (P/N F123615)
WITH D200 TIPS
10
± 0.35
≤
0.10
± 0.8
≤
0.3
50
± 0.40
≤
0.12
± 0.8
≤
0.3
100
± 0.80
≤
0.15
± 0.8
≤
0.3
P200 (P/N F123601)
WITH D200 TIPS
20
± 0.50
≤
0.20
± 1.6
≤
0.6
100
± 0.80
≤
0.25
± 1.6
≤
0.6
200
± 1.60
≤
0.30
± 1.6
≤
0.6
P1000 (P/N F123602)
WITH D1000 TIPS
100
± 3.0
≤
0.6
± 8.0
≤
3.0
500
± 4.0
≤
1.0
± 8.0
≤
3.0
1000
± 8.0
≤
1.5
± 8.0
≤
3.0
P5000* (P/N F123603)
WITH D5000 TIPS
500
± 12
≤
3
± 40
≤
15.0
2500
± 15
≤
5
± 40
≤
15.0
5000
± 30
≤
8
± 40
≤
15.0
P10mL* (P/N F161201)
WITH D10mL TIPS
1000
± 30
≤
6
± 60
≤
30.0
2000
± 30
≤
6
± 60
≤
30.0
5000
± 40
≤
10
± 60
≤
30.0
10000
± 60
≤
16
± 60
≤
30.0
*P5000 and P10mL do not have tip ejectors.
Summary of Contents for PIPETMAN Classic P10
Page 1: ...User s Guide EN CLASSIC ...
Page 25: ...Notes 25 PIPETMAN Classic USER S GUIDE NOTES ...
Page 26: ...Notes 26 Notes PIPETMAN Classic NOTES ...
Page 27: ...Notes 27 PIPETMAN Classic USER S GUIDE NOTES ...
Page 28: ...www gilson com contactus LT801120 H 2019 Gilson Inc All rights reserved ...