PIPETMAN® CLASSIC
|
USER’S GUIDE
TROUBLESHOOTING
|
PIPETMAN® CLASSIC
Chapter 6
MAINTENANCE
Routine maintenance will help keep your pipette
in good condition, ensuring a continued high
level of performance. Maintenance is limited to
cleaning or autoclaving the parts as specified
under
page 18 or to replacing the push button,
connecting nut, tip ejector, tip holder, seal, and
O-ring.
NOTICE
PIPETMAN P2 and P10 should
not be disassembled. You may
only replace the push button,
tip ejector, and its dual-position
adapter. With these pipettes, if
the tip holder is damaged, the
piston may also be damaged.
After replacing any parts you
should verify the performance
of your pipette following
the verification procedure
available on the Gilson website
). If the pipette
needs to be readjusted, please
contact your authorized Gilson
Service Center.
Changing the Tip Ejector
To remove:
Push the ejection
button.
Laterally push the tip
ejector.
Slide and remove the
tip ejector.
To refit:
Push the ejection
button.
Slide the tip ejector along the tip holder.
Clip the tip ejector on the body of the pipette.