TROUBLESHOOTING
|
PIPETMAN® CLASSIC
SYMPTOM
POSSIBLE CAUSE
ACTION
Pipette is not
precise
Tip holder is loose
Tighten connecting
nut
Connecting nut is
loose
Tighten connecting
nut
Incorrect operator
technique
Operator training
Damaged or corroded
piston
Return pipette to
supplier
Damaged tip holder
Replace the tip
holder
Worn O-ring or seal
Replace both parts
Tips fall off
or do not fit
correctly
Low quality tips
Use PIPETMAN
DIAMOND Tips
Damaged tip holder
Replace the tip
holder
Damaged tip ejector
Replace tip ejector
Ejector spacer is
missing
Mount the spacer on
the tip ejector
The ejector spacer is
damaged
Replace the ejector
spacer
The tip ejector is loose
Assemble the tip
ejector properly
The ejector lock is
misaligned
Align the ejector
lock
If you are not able to resolve the issue, contact
your Gilson representative.