16
USER’S GUIDE
|
ASPEC™ POSITIVE PRESSURE MANIFOLD
INTRODUCTION
Unpacking
The ASPEC® Positive Pressure Manifold is delivered as a base unit
(along with the safety shield) with one or more configuration kits for:
96-well SPE microplates (PN 37417014), 15-position SPE cartridge racks
(6 mL, PN 37417013), and 24-position SPE cartridge racks (1 mL, PN 37417011
or 3 mL, PN 37417012).
To safely remove the ASPEC PPM and accessories:
1.
Open the top of the shipping container.
2. Remove the packing slip and bag containing the documentation, which
will include: a printed user’s guide, USB drive with video, items included
checklist, and a quality control checklist.
3. Read the user’s guide.
4. Remove the box containing the standard equipment and accessory kit,
and then set it aside.
5. Check that all items specific to the configuration ordered and listed in
Standard Equipment
on page 17 are present.
6. Remove the safety shield nested in the foam, and then set it aside.
7. Grab the base of the ASPEC PPM firmly with two hands on opposing
sides, and gently lift the ASPEC PPM out of the shipping box, and
then set it on a laboratory bench that meets the site requirements on
page 19
.
NOTE
The ASPEC PPM ships and is nested within four molded
foam pieces: two on the front and two on the back.
Do not grip the ASPEC PPM by the manifold carrier or by
the levers.
8. Hold the base of the ASPEC PPM, and then gently slide the molded foam
pieces away from the base.