WindObserver 65
Page 15
Issue 4
Doc. No. 1390-PS-0039
June 2016
Default Settings
The factory default settings are:
Non - heated units
A0 B3 C1 E1 F1 G0000 J1 K1 L1 M2 NA O1 P1 T1 U1 V1 X1 Y1 Z1
Plus RWASHORT 10 and RWALONG 60.
Heated units
A0 B3 C1 E1 F1 G0000 H2 J1 K1 L1 M2 NA O1 P1 T1 U1 V1 X1 Y1 Z1
Plus RWASHORT 10 and RWALONG 60.
Note
that Heating (H command) is enabled as a default condition. If heating is not
required enabled then the H command must be set for H1.
See Section 8 Configuring for further details
Connecting to a PC with a RS422 or RS485 input
The PC requires either an internal RS422 interface card, or an external RS422/485 to
RS232 or USB converter, which will plug into the PC or host equipment.
WindObserver 65
Converter,
PC, or Logger
Signal
Names
19 Way
Connector
Cable Gland
RS422/RS485
Pin No.
Pair No
Colour
Signal Names
TXB (+)
P
1
Green
RXB (+)
TXA (-)
C
1
Black
RXA (-)
RXB (+)
U
2
White
TXA (+)
RXA (-)
V
2
Black
TXB (-)
Digital 0v
M
4
Blue
Signal Ground
Connected Internally
Ground Earth
R
3
Red
+ve Sensor
Sensor -
D
3
Black
-ve PSU
*
A
5
Yellow
+ve Heater
Heater -
B
5
Black
-ve PSU
*
An Ch 1
T
8
White
Ch 1 Device
An Ch 2
S
8
Red
Ch 2 Device
An Ch 3
H
6
Brown
Ch 3 Device
An Ch 0v
G
6
Black
Ch1-3 Return
*
For Power Supply Information see Section 6.4.