Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver 65
Page 47
Issue 4
Doc. No. 1390-PS-0039
June 2016
8.6.
Configuration Settings
Ax - Speed of Sound (SOS) and Temperature
Setting
Configuration code
Disable both
A0
Report SOS
A1
Report Sonic temperature
A2
Report SOS and Sonic temperature
A3
Note that neither of these parameters is available in Tunnel or NMEA formats.
Bx - Baud Rate
Setting
Configuration code
2400
B1
4800
B2
9600
B3
19200
B4
38400
B5
1200
B6
To change the Baud rate when using HyperTerminal: -
Eg. If set to B3 (9600 baud) and it is required to set to Baud rate to B2 (4800 baud).
Type * to enter Configuration Mode.
Type B 2 and press ENTER,
(Do not type any further commands at this stage).
Close the 9600 Baud HyperTerminal connection.
Open HyperTerminal and set the new connection Baud Rate to 4800 baud.
Type B and press ENTER, the letter B will be shown followed by B2
Type Q and press ENTER, data will scroll at the new baud rate.
Cx- Analogue Output Polar Mode Direction Wraparound
Setting
Configuration code
Wraparound (0 - 539
)
C1
Standard (0-360
)
C2