Ox – ASCII Output Format (Output String Padding)
Setting
Configuration code
Comma Separated Variable (CSV)
O1
Fixed Field
O2
Example data strings:-
POLAR
NMEA
CSV
data changing to error status code
condition.
CSV
data changing to error status code
condition
.
A,235,000.77,M,00
, 0A
$IIMWV,191,R,000.55,M,A*19
A,,,M,04, 24
$IIMWV,,R,,M,V*29
Fixed Field
data changing to error status
code condition.
Fixed Field
data changing to error status
code condition.
A,266,000.73,M,00, 08
$IIMWV,191,R,000.55,M,A*19
A,999,999.99,M,04, 0A
$IIMWV,999,R,999.99,M,V*07
Px - Output Rate
Outputs per second
1
2
4
5
8
10
Configuration code
P1
P3
P2
P4
P5
P6
Q- Measurement Mode
See Section 8.3.
RWASHORT XX (Short term number)
Where XX = 10 to 60, associated with Mode 14 and Mode 15 averaging.
RWALONG XX (Long term number)
Where XX = 1 to 10, associated with Mode 14 and Mode 15 averaging.
Tx - Analogue Output Voltage or Current settings (if fitted)
Setting
Configuration code
Voltage (0 to 5V)
T1
Voltage (–2.5V to +2.5V)
T2
Current (4-20mA)
T3
Ux – Digital Output Units
Units
metres/sec
(m/s)
knots
(knots)
miles / hour
(MPH)
kilometre/hour
(kph)
feet / minute
(fpm)
Configuration
code
U1
U2
U3
U4
U5