10

Signet 8250-2 Level Transmitter 

9. Options 

Menu

• 

The following entries are used by the 8250 to convert the level into volumetric and mass measurements. The accuracy of 
the entries entered will determine the accuracy of the volumetric and mass calculation. Refer to section 5 for additional 
information.

Display

(Factory settings shown)

Description

Contrast:

3 >

Adjust the LCD contrast for best viewing. A setting of 1 is lowest contrast, 5 is highest.  Select 
lower contrast if the display is in warmer ambient surroundings.

Display Option:

Level >

Select Level, Volume or Both.
This setting will determine the permanent display selection.

Averaging

Off >

OFF provides the quickest output response to changes in level or volume. Longer averaging 
period produces more stable display and output response.

Decimal

*****. >

Set the decimal to the best resolution for your application.  The display will automatically scale 
up to this resolution. Select *****., ****.*, ***.**, **.*** or *.****

Loop Adjust

4.00 mA 

>

Adjust the minimum and maximum current output. Use this setting to match the system output 
to any external device. The display value represents the precise current output.
Adjustment limits:
• 3.80 mA < 4.00 mA > 5.00 mA
• 19.00 mA < 20.00 mA > 21.00 mA

Loop Adjust

20.00 mA 

>

Test Loop:

 

>

Press UP or DOWN keys to manually order any output current value from 3.6 mA to 
21.00 mA to test current loop output.

Test Relay 1:

 

>

Press UP or DOWN keys to manually toggle the state of Relay 1.

Display

(Factory settings shown)

Description

VOL Shape:

Vert Cylinder 

>

Select the shape of the vessel where Sensor is located:

Vert Cylinder, Horiz Cylinder, Rectangular 

or 

Custom shape.

VOL Diameter:

0.0000 ft 

>

If Vert Cylinder or Horiz Cylinder is selected: 

Set the diameter of the cylinder.

The unit of measure is set by "LVL Units" in Calibrate menu.

VOL Length:

0.0000 ft 

>

If Horizontal Cylinder or Rectangular shape is selected: 

Set the length of the vessel.

The unit of measure is set by "LVL Units" in Calibrate menu.

VOL Width:

0.0000 ft 

>

If Rectangular shape is selected: 

Set the width of the vessel.

The unit of measure is set by "LVL Units" in Calibrate menu.

Number of Custom

Points: 10 

>

If Custom shape is selected: 

enter the number of measurement points to be used to de

fi

 ne 

the vessel shape. Minimum 3 points, maximum 10 points.

LVL Measurement:

Manual >

If Custom shape is selected: 

select Manual or Automatic level measurement.

Manual allows you to edit both the Level measurement and the Volume measurement.
Automatic allows you to edit only the Volume measurement (while displaying Level value)

P01: +0.0000 ft

V: +0.0000 gal  >

Set the Volume (and Level if manual measurement is selected) at each custom point in your 
vessel.

• 

Relay settings will repeat for Relay #2 and for External Relays.

• 

External Relays require the optional External Relay Module (model number 3-8059-4 or 3-8059-4AC).

• 

Note: When using the 8059-4 or -4AC, only relays 1 and 2 can be utilized with the fl ow instrument.

Summary of Contents for Signet 3-8250-2

Page 1: ...ble and reversible Max loop impedance 50 max 12 V 325 max 18 V 600 max 24 V Update rate 300 ms Output accuracy 0 03 mA Relay outputs 2 sets mechanical SPDT contacts Software supports 2 additional rela...

Page 2: ...ure instrument temporarily with tape from front or grip from rear of instrument DO NOT RELEASE 5 Press quick clips outward and remove 2 1 Terminals 11 13 Digital S3 L The 2450 sensor and External Rela...

Page 3: ...el 1 4 to 20 mA in Channel 2 4 to 20 mA in System Pwr Loop System Pwr Loop AUX Power AUX Power System Pwr Loop System Pwr Loop AUX Power AUX Power Loop Input 4 to 20 mA System Pwr Loop System Pwr Loop...

Page 4: ...menu if the sensor is located at Z Level L The distance from Z to surface of fluid displayed as Level by transmitter L 0 O 0 Z SLOC L 0 O 0 Z SLOC L 0 SLOC O 0 Z 2450 Pressure Sensor Location and Offs...

Page 5: ...ted at each transition point in the vessel shape for example at custom point 9 where the shape changes from a cylinder to a cone The more complex sections should be defined with more points Notice tha...

Page 6: ...to change display selection Output settings cannot be edited from the VIEW menu When using the CALIBRATE or OPTIONS menus the ProcessPro will return to the VIEW menu if no key is pressed for 10 minut...

Page 7: ...g Procedure Step 1 Press and hold ENTER key 2 seconds to select the CALIBRATE menu 5 seconds to select the OPTIONS menu Step 2 The Key Code is UP UP UP DOWN keys in sequence After entering the Key Cod...

Page 8: ...L Single point calibration for Level Reset LVL to Factory Cal Yes Reset Level to original factory calibration Set LVL entry is erased No Return to menu do not reset Level Loop Source LVL Select the so...

Page 9: ...for relay response Relay will be activated only if the source value exceeds the setpoint for this time period Relay settings will repeat for Relay 2 and for External Relays External Relays require th...

Page 10: ...t Test Relay 1 Press UP or DOWN keys to manually toggle the state of Relay 1 Display Factory settings shown Description VOL Shape Vert Cylinder Select the shape of the vessel where Sensor is located V...

Page 11: ...iew all custom points and reset values so each point is greater than the preceding point CHK SENSOR Problem with the sensor during different modes of operation 1 Wiring problem AUXILIARY POWER was not...

Page 12: ...upply Note When using the 8059 4 or 8059 4AC only relays 1 and 2 can be utilized with the flow instrument Accessories Mfr Part No Code Description 3 8050 159 000 184 Universal mounting kit 3 8050 395...

Reviews: